Two-photon excitation microscopy (TPEF or 2PEF) is a
fluorescence
Fluorescence is one of two kinds of photoluminescence, the emission of light by a substance that has absorbed light or other electromagnetic radiation. When exposed to ultraviolet radiation, many substances will glow (fluoresce) with colore ...
imaging technique that is particularly well-suited to image scattering
living tissue of up to about one millimeter in thickness. Unlike traditional
fluorescence microscopy, where the excitation wavelength is shorter than the emission wavelength,
two-photon excitation requires simultaneous excitation by two photons with longer wavelength than the emitted light. The laser is focused onto a specific location in the tissue and scanned across the sample to sequentially produce the image. Due to the
non-linearity of two-photon excitation, mainly fluorophores in the micrometer-sized focus of the laser beam are excited, which results in the spatial resolution of the image. This contrasts with
confocal microscopy, where the spatial resolution is produced by the interaction of excitation focus and the confined detection with a pinhole.
Two-photon excitation microscopy typically uses
near-infrared
Infrared (IR; sometimes called infrared light) is electromagnetic radiation (EMR) with wavelengths longer than that of visible light but shorter than microwaves. The infrared spectral band begins with the waves that are just longer than those of ...
(NIR) excitation light which can also excite
fluorescent dyes. Using infrared light minimizes scattering in the tissue because infrared light is scattered less in typical biological tissues. Due to the multiphoton absorption, the background signal is strongly suppressed. Both effects lead to an increased
penetration depth for this technique. Two-photon excitation can be a superior alternative to
confocal microscopy due to its deeper tissue penetration, efficient light detection, and reduced
photobleaching.
upright=1.5, Two-photon fluorescence image (green) of a cross section of colored with lily of the valley">rhizome colored with lily of the valley. The excitement is at 840 nm, and the red and blue colors represent other channels of multiphoton techniques which have been superimposed.
Concept
Two-photon excitation employs
two-photon absorption, a concept first described by Maria Goeppert Mayer (1906–1972) in her doctoral dissertation in 1931,
[
] and first observed in 1961 in a CaF
2:Eu
2+ crystal using
laser
A laser is a device that emits light through a process of optical amplification based on the stimulated emission of electromagnetic radiation. The word ''laser'' originated as an acronym for light amplification by stimulated emission of radi ...
excitation by
Wolfgang Kaiser.
Isaac Abella showed in 1962 in
caesium
Caesium (IUPAC spelling; also spelled cesium in American English) is a chemical element; it has Symbol (chemistry), symbol Cs and atomic number 55. It is a soft, silvery-golden alkali metal with a melting point of , which makes it one of only f ...
vapor that two-photon excitation of single atoms is possible.
Two-photon excitation fluorescence microscopy has similarities to other confocal laser microscopy techniques such as
laser scanning confocal microscopy and
Raman microscopy
The Raman microscope is a laser-based microscope, microscopic device used to perform Raman spectroscopy.''Microscopical techniques in the use of the molecular optics laser examiner Raman microprobe'', by M. E. Andersen, R. Z. Muggli, Analytical C ...
. These techniques use focused laser beams scanned in a raster pattern to generate images, and both have an
optical sectioning effect. Unlike confocal microscopes, multiphoton microscopes do not contain pinhole apertures that give confocal microscopes their optical sectioning quality. The optical sectioning produced by multiphoton microscopes is a result of the
point spread function of the excitation.
The concept of two-photon excitation is based on the idea that two photons, of comparably lower photon energy than needed for one-photon excitation, can also excite a
fluorophore in one quantum event. Each photon carries approximately half the energy necessary to excite the molecule. The emitted photon is at a higher energy (shorter wavelength) than either of the two exciting photons. The probability of the near-simultaneous absorption of two photons is extremely low. Therefore, a high peak
flux of excitation photons is typically required, usually generated by femtosecond
pulsed laser. For example, the same average laser power but without pulsing results in no detectable fluorescence compared to fluorescence generated by the pulsed laser via the two-photon effect.
The longer wavelength, lower energy (typically infrared) excitation lasers of multiphoton microscopes are well-suited to use in imaging live cells as they cause less damage than the short-wavelength lasers typically used for single-photon excitation, so living tissues may be observed for longer periods with fewer toxic effects.
The most commonly used fluorophores have excitation spectra in the 400–500 nm range, whereas the laser used to excite the two-photon fluorescence lies in the ~700–1100 nm (infrared) range produced by
Ti-sapphire lasers. If the fluorophore absorbs two infrared photons simultaneously, it will absorb enough energy to be raised into the excited state. The fluorophore will then emit a single photon with a wavelength that depends on the type of fluorophore used (typically in the
visible spectrum
The visible spectrum is the spectral band, band of the electromagnetic spectrum that is visual perception, visible to the human eye. Electromagnetic radiation in this range of wavelengths is called ''visible light'' (or simply light).
The optica ...
). Because two photons are absorbed during the excitation of the fluorophore, the probability of fluorescent emission from the fluorophores increases quadratically with the excitation intensity. Therefore, much more two-photon fluorescence is generated where the laser beam is tightly focused than where it is more diffuse. Effectively, excitation is restricted to the tiny focal volume (~1 femtoliter), resulting in a high degree of rejection of out-of-focus objects. This ''localization of excitation'' is the key advantage compared to
single-photon excitation microscopes, which need to employ elements such as pinholes to reject out-of-focus fluorescence. The fluorescence from the sample is then collected by a high-sensitivity detector, such as a
photomultiplier tube. This observed light intensity becomes one
pixel
In digital imaging, a pixel (abbreviated px), pel, or picture element is the smallest addressable element in a Raster graphics, raster image, or the smallest addressable element in a dot matrix display device. In most digital display devices, p ...
in the eventual image; the focal point is scanned throughout a desired region of the sample to form all the pixels of the image.
Development
Two-photon microscopy was pioneered and
patent
A patent is a type of intellectual property that gives its owner the legal right to exclude others from making, using, or selling an invention for a limited period of time in exchange for publishing an sufficiency of disclosure, enabling discl ...
ed by
Winfried Denk and James Strickler in the lab of
Watt W. Webb at
Cornell University
Cornell University is a Private university, private Ivy League research university based in Ithaca, New York, United States. The university was co-founded by American philanthropist Ezra Cornell and historian and educator Andrew Dickson W ...
in 1990. They combined the idea of
two-photon absorption with the use of a laser scanner.
In two-photon excitation microscopy an infrared laser beam is focused through an objective lens. The
Ti-sapphire laser normally used has a pulse width of approximately 100 femtoseconds (fs) and a repetition rate of about 80
MHz, allowing the high photon density and flux required for two-photon absorption, and is tunable across a wide range of wavelengths.
The use of infrared light to excite fluorophores in
light-scattering tissue has added benefits.
Longer wavelengths are scattered to a lesser degree than shorter ones, which is a benefit to high-resolution imaging. In addition, these lower-energy photons are less likely to cause damage outside the focal volume. Compared to a confocal microscope, photon detection is much more effective since even scattered photons contribute to the usable signal. These benefits for imaging in scattering tissues were only recognized several years after the invention of two-photon excitation microscopy.
There are several caveats to using two-photon microscopy: The pulsed lasers needed for two-photon excitation are much more expensive than the continuous wave (CW) lasers used in confocal microscopy. The two-photon absorption spectrum of a molecule may vary significantly from its one-photon counterpart. Higher-order photodamage becomes a problem and bleaching scales with the square of the laser power, whereas it is linear for single-photon (confocal). For very thin objects such as isolated cells, single-photon (confocal) microscopes can produce images with higher
optical resolution due to their shorter excitation wavelengths. In scattering tissue, on the other hand, the superior optical sectioning and light detection capabilities of the two-photon microscope result in better performance.
Applications
Main
Two-photon microscopy has been involved in numerous fields including: physiology, neurobiology, embryology and
tissue engineering. Even thin, nearly transparent tissues (such as skin cells) have been visualized with clear detail due to this technique.
Two-photon microscopy's high speed imaging capabilities may also be utilized in noninvasive optical biopsy.
Two-photon microscopy has been aptly used for producing localized chemical reactions,
an effect that has been used also for
two-photon-based lithography. Using two-photon fluorescence and
second-harmonic generation–based microscopy, it was shown that organic
porphyrin-type molecules can have different
transition dipole moments for two-photon fluorescence and second harmonic generation,
which are otherwise thought to occur from the same transition dipole moment.
Non-degenerative two-photon excitation, or using 2 photons of unequal wavelengths, was shown to increase the fluorescence of all tested small molecules and fluorescent proteins.
Cancer research
2PEF was also proven to be very valuable for characterizing
skin cancer,
It had also been shown to reveal tumor cell arrest, tumor cell-platelet interaction, tumor cell-leukocyte interaction and metastatic colonization processes.
Embryonic research
2PEF has shown to be advantageous over other techniques, such as
confocal microscopy when it comes to long-term live-cell imaging of mammalian embryos.
Kidney research
2PEF has also been used in visualization of difficult-to-access cell types, especially in regards to kidney cells. It has been used in better understanding fluid dynamics and filtration.
Viral infection level determination
2PEF was also proven to be valuable tool for monitoring correlates of viral (
SARS-CoV-2
Severe acute respiratory syndrome coronavirus 2 (SARS‑CoV‑2) is a strain of coronavirus that causes COVID-19, the respiratory illness responsible for the COVID-19 pandemic. The virus previously had the Novel coronavirus, provisional nam ...
) infection in cell culture using a 2P-active Ca
2+ sensitive dye.
Neuroscience
2PEF as well as the extension of this method to
3PEF are used to characterize intact neural tissues in the brain of living and even behaving animals. In particular, the method is advantageous for
calcium imaging of a neuron or populations of neurons,
for
photopharmacology including localized uncaging of components such as glutamate,
GABA or
isomerization of photoswitchable drugs, and for the imaging of other genetically encoded sensors that report the concentration of neurotransmitters.
Currently, two-photon microscopy is widely used to image the live firing of neurons in model organisms including fruit flies (''
Drosophila melanogaster)'',
rats,
songbirds,
primates,
ferrets, mice (''
Mus musculus
The house mouse (''Mus musculus'') is a small mammal of the rodent family Muridae, characteristically having a pointed snout, large rounded ears, and a long and almost hairless tail. It is one of the most abundant species of the genus ''Mus (genu ...
)'',
zebrafish. The animals are typically head-fixed due to the size of the microscope and scan devices, but also miniatured microscopes are being developed that enable imaging of neurons in the moving and freely behaving animals.
[ ]
Higher-order excitation
Simultaneous absorption of three or more photons is also possible, allowing for higher-order multiphoton excitation microscopy.
So-called "three-photon excitation fluorescence microscopy" (3PEF) is the most used technique after 2PEF, to which it is complementary. Localized
isomerization of photoswitchable drugs in vivo using three-photon excitation has also been reported.
Dyes and fluorescent proteins for two-photon excitation microscopy
In general, all commonly used
fluorescent proteins (CFP, GFP, YFP, RFP) and dyes can be excited in two-photon mode. Two-photon excitation spectra are often considerably broader, making it more difficult to excite fluorophores selectively by switching excitation wavelengths.
Several green, red and NIR emitting dyes (probes and reactive labels) with extremely high 2-photon absorption cross-sections have been reported.
Due to the donor-acceptor-donor type structure,
squaraine dyes such as ''Seta-670'', ''Seta-700'' and ''Seta-660'' exhibit very high 2-photon absorption (2PA) efficiencies in comparison to other dyes,
''SeTau-647'' and ''SeTau-665'', a new type of squaraine-
rotaxane, exhibit extremely high two-photon action cross-sections of up to 10,000 GM in the near IR region, unsurpassed by any other class of organic dyes.
See also
*
3D optical data storage
*
Nonlinear optics
*
Second-harmonic imaging microscopy
*
Three-photon microscopy
*
Two-photon absorption
*
Two-photon photoelectron spectroscopy
*
Wide-field multiphoton microscopy
Sources
*
*
*
*
References
External links
Simplifying two-photon microscopyWebinar: Setting Up a Simple and Cost-Efficient 2Photon MicroscopeTwo-photon suitable dyesAcquisition of Multiple Real-Time Images for Laser Scanning Microscopy(Sanderson microscopy article)
Two-photon Fluorescence Light Microscopy, ENCYCLOPEDIA OF LIFE SCIENCES*
University of Wisconsin.
*
'' Nikon MicroscopyU .
*
*
{{Lasers
Microscopy
Cell imaging
Fluorescence techniques
Optical microscopy