Single-photon emission computed tomography (SPECT, or less commonly, SPET) is a nuclear medicine tomographic imaging technique using gamma rays. It is very similar to conventional nuclear medicine planar imaging using a gamma camera (that is, scintigraphy), but is able to provide true 3D information. This information is typically presented as cross-sectional slices through the patient, but can be freely reformatted or manipulated as required. The technique needs delivery of a gamma-emitting radioisotope (a
radionuclide A radionuclide (radioactive nuclide, radioisotope or radioactive isotope) is a nuclide that has excess nuclear energy, making it unstable. This excess energy can be used in one of three ways: emitted from the nucleus as gamma radiation; transfer ...
) into the patient, normally through injection into the bloodstream. On occasion, the radioisotope is a simple soluble dissolved ion, such as an isotope of gallium(III). Most of the time, though, a marker radioisotope is attached to a specific ligand to create a radioligand, whose properties bind it to certain types of tissues. This marriage allows the combination of ligand and radiopharmaceutical to be carried and bound to a place of interest in the body, where the ligand concentration is seen by a gamma camera.


Principles

Instead of just "taking a picture of anatomical structures", a SPECT scan monitors level of biological activity at each place in the 3-D region analyzed. Emissions from the radionuclide indicate amounts of blood flow in the capillaries of the imaged regions. In the same way that a plain X-ray is a 2-dimensional (2-D) view of a 3-dimensional structure, the image obtained by a gamma camera is a 2-D view of 3-D distribution of a
radionuclide A radionuclide (radioactive nuclide, radioisotope or radioactive isotope) is a nuclide that has excess nuclear energy, making it unstable. This excess energy can be used in one of three ways: emitted from the nucleus as gamma radiation; transfer ...
. SPECT imaging is performed by using a gamma camera to acquire multiple 2-D images (also called projections), from multiple angles. A computer is then used to apply a tomographic reconstruction algorithm to the multiple projections, yielding a 3-D data set. This data set may then be manipulated to show thin slices along any chosen axis of the body, similar to those obtained from other tomographic techniques, such as
magnetic resonance imaging Magnetic resonance imaging (MRI) is a medical imaging technique used in radiology to form pictures of the anatomy and the physiological processes of the body. MRI scanners use strong magnetic fields, magnetic field gradients, and radio wave ...
(MRI), X-ray computed tomography (X-ray CT), and
positron emission tomography Positron emission tomography (PET) is a functional imaging technique that uses radioactive substances known as radiotracers to visualize and measure changes in Metabolism, metabolic processes, and in other physiological activities including bl ...
(PET). SPECT is similar to PET in its use of radioactive tracer material and detection of gamma rays. In contrast with PET, the tracers used in SPECT emit gamma radiation that is measured directly, whereas PET tracers emit positrons that annihilate with electrons up to a few millimeters away, causing two gamma photons to be emitted in opposite directions. A PET scanner detects these emissions "coincident" in time, which provides more radiation event localization information and, thus, higher spatial resolution images than SPECT (which has about 1 cm resolution). SPECT scans are significantly less expensive than PET scans, in part because they are able to use longer-lived and more easily obtained radioisotopes than PET. Because SPECT acquisition is very similar to planar gamma camera imaging, the same radiopharmaceuticals may be used. If a patient is examined in another type of nuclear medicine scan, but the images are non-diagnostic, it may be possible to proceed straight to SPECT by moving the patient to a SPECT instrument, or even by simply reconfiguring the camera for SPECT image acquisition while the patient remains on the table. To acquire SPECT images, the gamma camera is rotated around the patient. Projections are acquired at defined points during the rotation, typically every 3–6 degrees. In most cases, a full 360-degree rotation is used to obtain an optimal reconstruction. The time taken to obtain each projection is also variable, but 15–20 seconds is typical. This gives a total scan time of 15–20 minutes. Multi-headed gamma cameras can accelerate acquisition. For example, a dual-headed camera can be used with heads spaced 180 degrees apart, allowing two projections to be acquired simultaneously, with each head requiring 180 degrees of rotation. Triple-head cameras with 120-degree spacing are also used. Cardiac gated acquisitions are possible with SPECT, just as with planar imaging techniques such as multi gated acquisition scan (MUGA). Triggered by electrocardiogram (EKG) to obtain differential information about the heart in various parts of its cycle, gated myocardial SPECT can be used to obtain quantitative information about myocardial perfusion, thickness, and contractility of the myocardium during various parts of the cardiac cycle, and also to allow calculation of
left ventricular ejection fraction An ejection fraction (EF) is the volumetric fraction (or portion of the total) of fluid (usually blood) ejected from a chamber (usually the heart) with each contraction (or heartbeat). It can refer to the cardiac atrium, ventricle, gall bladder, o ...
, stroke volume, and cardiac output.


Application

SPECT can be used to complement any gamma imaging study, where a true 3D representation can be helpful, such as tumor imaging, infection (
leukocyte White blood cells, also called leukocytes or leucocytes, are the cells of the immune system that are involved in protecting the body against both infectious disease and foreign invaders. All white blood cells are produced and derived from mult ...
) imaging, thyroid imaging or
bone scintigraphy A bone scan or bone scintigraphy is a nuclear medicine imaging technique of the bone. It can help diagnose a number of bone conditions, including cancer of the bone or metastasis, location of bone inflammation and fractures (that may not be visi ...
. Because SPECT permits accurate localisation in 3D space, it can be used to provide information about localised function in internal organs, such as functional cardiac or brain imaging.


Myocardial perfusion imaging

Myocardial perfusion imaging (MPI) is a form of functional cardiac imaging, used for the diagnosis of
ischemic heart disease Coronary artery disease (CAD), also called coronary heart disease (CHD), ischemic heart disease (IHD), myocardial ischemia, or simply heart disease, involves the reduction of blood flow to the heart muscle due to build-up of atherosclerotic pla ...
. The underlying principle is that under conditions of stress, diseased
myocardium Cardiac muscle (also called heart muscle, myocardium, cardiomyocytes and cardiac myocytes) is one of three types of vertebrate muscle tissues, with the other two being skeletal muscle and smooth muscle. It is an involuntary, striated muscle that ...
receives less blood flow than normal myocardium. MPI is one of several types of
cardiac stress test A cardiac stress test (also referred to as a cardiac diagnostic test, cardiopulmonary exercise test, or abbreviated CPX test) is a cardiological test that measures the heart's ability to respond to external stress in a controlled clinical environ ...
. A cardiac specific radiopharmaceutical is administered, e.g., 99mTc-
tetrofosmin Technetium (99mTc) tetrofosmin is a drug used in nuclear medicine cardiac imaging. It is sold under the brand name Myoview (GE Healthcare). The radioisotope, technetium-99m, is chelated by two 1,2-''bis''i-(2-ethoxyethyl)phosphinothane ligands whi ...
(Myoview, GE healthcare), 99m Tc-sestamibi (Cardiolite, Bristol-Myers Squibb) or Thallium-201 chloride. Following this, the heart rate is raised to induce myocardial stress, either by exercise on a treadmill or pharmacologically with
adenosine Adenosine ( symbol A) is an organic compound that occurs widely in nature in the form of diverse derivatives. The molecule consists of an adenine attached to a ribose via a β-N9-glycosidic bond. Adenosine is one of the four nucleoside building ...
, dobutamine, or dipyridamole (
aminophylline Aminophylline is a compound of the bronchodilator theophylline with ethylenediamine in 2:1 ratio. The ethylenediamine improves solubility, and the aminophylline is usually found as a dihydrate. Aminophylline is less potent and shorter-acting tha ...
can be used to reverse the effects of dipyridamole). SPECT imaging performed after stress reveals the distribution of the radiopharmaceutical, and therefore the relative blood flow to the different regions of the myocardium. Diagnosis is made by comparing stress images to a further set of images obtained at rest which are normally acquired prior to the stress images. MPI has been demonstrated to have an overall accuracy of about 83% (
sensitivity Sensitivity may refer to: Science and technology Natural sciences * Sensitivity (physiology), the ability of an organism or organ to respond to external stimuli ** Sensory processing sensitivity in humans * Sensitivity and specificity, statisti ...
: 85%; specificity: 72%) (in a review, not exclusively of SPECT MPI), and is comparable with (or better than) other non-invasive tests for ischemic heart disease.


Functional brain imaging

Usually, the gamma-emitting tracer used in functional brain imaging is Technetium (99mTc) exametazime. 99mTc is a metastable nuclear isomer that emits gamma rays detectable by a gamma camera. Attaching it to exametazime allows it to be taken up by brain tissue in a manner proportional to brain blood flow, in turn allowing cerebral blood flow to be assessed with the nuclear gamma camera. Because blood flow in the brain is tightly coupled to local brain metabolism and energy use, the 99mTc-exametazime tracer (as well as the similar 99mTc-EC tracer) is used to assess brain metabolism regionally, in an attempt to diagnose and differentiate the different causal pathologies of dementia. Meta-analysis of many reported studies suggests that SPECT with this tracer is about 74% sensitive at diagnosing Alzheimer's disease vs. 81% sensitivity for clinical exam ( cognitive testing, etc.). More recent studies have shown the accuracy of SPECT in Alzheimer's diagnosis may be as high as 88%. In meta analysis, SPECT was superior to clinical exam and clinical criteria (91% vs. 70%) in being able to differentiate Alzheimer's disease from vascular dementias. This latter ability relates to SPECT's imaging of local metabolism of the brain, in which the patchy loss of cortical metabolism seen in multiple strokes differs clearly from the more even or "smooth" loss of non-occipital cortical brain function typical of Alzheimer's disease. Another recent review article showed that multi-headed SPECT cameras with quantitative analysis result in an overall sensitivity of 84-89% and an overall specificity of 83-89% in cross sectional studies and sensitivity of 82-96% and specificity of 83-89% for longitudinal studies of dementia. 99mTc-exametazime SPECT scanning competes with fludeoxyglucose (FDG) PET scanning of the brain, which works to assess regional brain glucose metabolism, to provide very similar information about local brain damage from many processes. SPECT is more widely available, because the radioisotope used is longer-lasting and far less expensive in SPECT, and the gamma scanning equipment is less expensive as well. While 99mTc is extracted from relatively simple technetium-99m generators, which are delivered to hospitals and scanning centers weekly to supply fresh radioisotope, FDG PET relies on FDG, which is made in an expensive medical cyclotron and "hot-lab" (automated chemistry lab for radiopharmaceutical manufacture), and then delivered immediately to scanning sites because of the natural short 110-minute half-life of Fluorine-18.


Applications in nuclear technology

In the nuclear power sector, the SPECT technique can be applied to image radioisotope distributions in irradiated nuclear fuels. Due to the irradiation of nuclear fuel (e.g. uranium) with neutrons in a nuclear reactor, a wide array of gamma-emitting radionuclides are naturally produced in the fuel, such as fission products ( cesium-137, barium-140 and europium-154) and activation products (
chromium-51 Naturally occurring chromium (24Cr) is composed of four stable isotopes; 50Cr, 52Cr, 53Cr, and 54Cr with 52Cr being the most abundant (83.789% natural abundance). 50Cr is suspected of decaying by β+β+ to 50Ti with a half-life of (more than) ...
and cobalt-58). These may be imaged using SPECT in order to verify the presence of fuel rods in a stored fuel assembly for IAEA safeguards purposes, to validate predictions of core simulation codes, or to study the behavior of the nuclear fuel in normal operation, or in accident scenarios.


Reconstruction

Reconstructed images typically have resolutions of 64×64 or 128×128 pixels, with the pixel sizes ranging from 3–6 mm. The number of projections acquired is chosen to be approximately equal to the width of the resulting images. In general, the resulting reconstructed images will be of lower resolution, have increased noise than planar images, and be susceptible to artifacts. Scanning is time-consuming, and it is essential that there is no patient movement during the scan time. Movement can cause significant degradation of the reconstructed images, although movement compensation reconstruction techniques can help with this. A highly uneven distribution of radiopharmaceutical also has the potential to cause artifacts. A very intense area of activity (e.g., the bladder) can cause extensive streaking of the images and obscure neighboring areas of activity. This is a limitation of the
filtered back projection In mathematics, the Radon transform is the integral transform which takes a function ''f'' defined on the plane to a function ''Rf'' defined on the (two-dimensional) space of lines in the plane, whose value at a particular line is equal to the ...
reconstruction algorithm. Iterative reconstruction is an alternative algorithm that is growing in importance, as it is less sensitive to artifacts and can also correct for attenuation and depth dependent blurring. Furthermore, iterative algorithms can be made more efficacious using the
Superiorization Superiorization is an iterative method for constrained optimization. It is used for improving the efficacy of an iterative method whose convergence is resilient to certain kinds of perturbations. Such perturbations are designed to "force" the per ...
methodology. Attenuation of the gamma rays within the patient can lead to significant underestimation of activity in deep tissues, compared to superficial tissues. Approximate correction is possible, based on relative position of the activity, and optimal correction is obtained with measured attenuation values. Modern SPECT equipment is available with an integrated X-ray CT scanner. As X-ray CT images are an attenuation map of the tissues, this data can be incorporated into the SPECT reconstruction to correct for attenuation. It also provides a precisely
registered Registered may refer to: * Registered mail, letters, packets or other postal documents considered valuable and in need of a chain of custody * Registered trademark symbol, symbol ® that provides notice that the preceding is a trademark or service ...
CT image, which can provide additional anatomical information. Scatter of the gamma rays as well as the random nature of gamma rays can also lead to the degradation of quality of SPECT images and cause loss of resolution. Scatter correction and resolution recovery are also applied to improve resolution of SPECT images.


Typical SPECT acquisition protocols


SPECT/CT

In some cases a SPECT gamma scanner may be built to operate with a conventional CT scanner, with coregistration of images. As in
PET/CT Positron emission tomography–computed tomography (better known as PET-CT or PET/CT) is a nuclear medicine technique which combines, in a single gantry, a positron emission tomography (PET) scanner and an x-ray computed tomography (CT) scanner ...
, this allows location of tumors or tissues which may be seen on SPECT scintigraphy, but are difficult to locate precisely with regard to other anatomical structures. Such scans are most useful for tissues outside the brain, where location of tissues may be far more variable. For example, SPECT/CT may be used in
sestamibi parathyroid scan A sestamibi parathyroid scan is a procedure in nuclear medicine which is performed to localize parathyroid adenoma, which causes Hyperparathyroidism. Adequate localization of parathyroid adenoma allows the surgeon to use a minimally invasive surgi ...
applications, where the technique is useful in locating ectopic parathyroid adenomas which may not be in their usual locations in the thyroid gland.


Quality control

The overall performance of SPECT systems can be performed by quality control tools such as the Jaszczak phantom.Jennifer Prekeges. ''Nuclear Medicine Instrumentation''. Jones & Bartlett Publishers. 2012. p.189


See also

* Daniel Amen, psychiatrist who uses SPECT for diagnoses * Functional neuroimaging * Gamma camera *
Magnetic resonance imaging Magnetic resonance imaging (MRI) is a medical imaging technique used in radiology to form pictures of the anatomy and the physiological processes of the body. MRI scanners use strong magnetic fields, magnetic field gradients, and radio wave ...
* Neuroimaging *
Positron emission tomography Positron emission tomography (PET) is a functional imaging technique that uses radioactive substances known as radiotracers to visualize and measure changes in Metabolism, metabolic processes, and in other physiological activities including bl ...
* ISAS (Ictal-Interictal SPECT Analysis by SPM)


References

*


Further reading

* Bruyant, P. P. (2002)
"Analytic and iterative reconstruction algorithms in SPECT"
''Journal of Nuclear Medicine'' 43(10):1343-1358. * Elhendy et al.
"Dobutamine Stress Myocardial Perfusion Imaging in Coronary Artery Disease"
''J Nucl Med'' 2002 43: 1634–1646. * * * Jones / Hogg / Seeram (2013). ''Practical SPECT/CT in Nuclear Medicine''. . * Willowson K, Bailey DL, Baldock C, 2008. "Quantitative SPECT reconstruction using CT-derived corrections". ''Phys. Med. Biol.'' 53 3099–3112.


External links


Human Health Campus, The official website of the International Atomic Energy Agency dedicated to Professionals in Radiation Medicine. This site is managed by the Division of Human Health, Department of Nuclear Sciences and ApplicationsNational Isotope Development Center
Reference information on radioisotopes including those for SPECT; coordination and management of isotope production, availability, and distribution
Isotope Development & Production for Research and Applications (IDPRA)
U.S. Department of Energy program for isotope production and production research and development {{DEFAULTSORT:Single Photon Emission Computed Tomography 3D nuclear medical imaging Radiobiology Neuroimaging Medical physics Articles containing video clips