Micellar liquid chromatography (MLC) is a form of reversed phase
liquid chromatography In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent (gas or liquid) called the ''mobile phase'', which carries it through a system ( ...
that uses an
aqueous An aqueous solution is a solution in which the solvent is water. It is mostly shown in chemical equations by appending (aq) to the relevant chemical formula. For example, a solution of table salt, or sodium chloride (NaCl), in water would be ...
micellar solutions as the mobile phase.


Theory

The use of micelles in high performance liquid chromatography was first introduced by Armstrong and Henry in 1980. The technique is used mainly to enhance retention and selectivity of various
solutes In chemistry, a solution is a special type of homogeneous mixture composed of two or more substances. In such a mixture, a solute is a substance dissolved in another substance, known as a solvent. If the attractive forces between the solvent ...
that would otherwise be inseparable or poorly resolved. Micellar liquid chromatography (MLC) has been used in a variety of applications including separation of
mixture In chemistry, a mixture is a material made up of two or more different chemical substances which are not chemically bonded. A mixture is the physical combination of two or more substances in which the identities are retained and are mixed in the ...
s of charged and neutral solutes, direct injection of serum and other physiological fluids, analysis of pharmaceutical compounds, separation of
enantiomer In chemistry, an enantiomer ( /ɪˈnænti.əmər, ɛ-, -oʊ-/ ''ih-NAN-tee-ə-mər''; from Ancient Greek ἐνάντιος ''(enántios)'' 'opposite', and μέρος ''(méros)'' 'part') – also called optical isomer, antipode, or optical ant ...
s, analysis of inorganic
organometallic Organometallic chemistry is the study of organometallic compounds, chemical compounds containing at least one chemical bond between a carbon atom of an organic molecule and a metal, including alkali, alkaline earth, and transition metals, and ...
s, and a host of others. One of the main drawbacks of the technique is the reduced efficiency that is caused by the micelles. Despite the sometimes poor efficiency, MLC is a better choice than ion-exchange LC or ion-pairing LC for separation of charged
molecule A molecule is a group of two or more atoms held together by attractive forces known as chemical bonds; depending on context, the term may or may not include ions which satisfy this criterion. In quantum physics, organic chemistry, and bioch ...
s and mixtures of charged and neutral
species In biology, a species is the basic unit of Taxonomy (biology), classification and a taxonomic rank of an organism, as well as a unit of biodiversity. A species is often defined as the largest group of organisms in which any two individuals of ...
. Some of the aspects which will be discussed are the theoretical aspects of MLC, the use of models in predicting retentive characteristics of MLC, the effect of micelles on efficiency and selectivity, and general applications of MLC. Reverse phase
high-performance liquid chromatography High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify each component in a mixture. It relies on pumps to p ...
(RP-HPLC) involves a non- polar stationary phase, often a hydrocarbon chain, and a polar mobile or liquid phase. The mobile phase generally consists of an
aqueous An aqueous solution is a solution in which the solvent is water. It is mostly shown in chemical equations by appending (aq) to the relevant chemical formula. For example, a solution of table salt, or sodium chloride (NaCl), in water would be ...
portion with an organic addition, such as
methanol Methanol (also called methyl alcohol and wood spirit, amongst other names) is an organic chemical and the simplest aliphatic alcohol, with the formula C H3 O H (a methyl group linked to a hydroxyl group, often abbreviated as MeOH). It is a lig ...
or
acetonitrile Acetonitrile, often abbreviated MeCN (methyl cyanide), is the chemical compound with the formula and structure . This colourless liquid is the simplest organic nitrile (hydrogen cyanide is a simpler nitrile, but the cyanide anion is not classed ...
. When a solution of
analyte An analyte, component (in clinical chemistry), or chemical species is a substance or chemical constituent that is of interest in an analytical procedure. The purest substances are referred to as analytes, such as 24 karat gold, NaCl, water, e ...
s is injected into the system, the components begin to partition out of the mobile phase and interact with the stationary phase. Each component interacts with the stationary phase in a different manner depending upon its polarity and
hydrophobicity In chemistry, hydrophobicity is the physical property of a molecule that is seemingly repelled from a mass of water (known as a hydrophobe). In contrast, hydrophiles are attracted to water. Hydrophobic molecules tend to be nonpolar and, t ...
. In reverse phase HPLC, the solute with the greatest polarity will interact less with the stationary phase and spend more time in the mobile phase. As the polarity of the components decreases, the time spent in the column increases. Thus, a separation of components is achieved based on polarity. The addition of micelles to the mobile phase introduces a third phase into which the solutes may partition.


Micelles

Micelle A micelle () or micella () (plural micelles or micellae, respectively) is an aggregate (or supramolecular assembly) of surfactant amphipathic lipid molecules dispersed in a liquid, forming a colloidal suspension (also known as associated coll ...
s are composed of surfactant, or detergent,
monomer In chemistry, a monomer ( ; ''mono-'', "one" + ''-mer'', "part") is a molecule that can react together with other monomer molecules to form a larger polymer chain or three-dimensional network in a process called polymerization. Classification M ...
s with a hydrophobic
moiety Moiety may refer to: Chemistry * Moiety (chemistry), a part or functional group of a molecule ** Moiety conservation, conservation of a subgroup in a chemical species Anthropology * Moiety (kinship), either of two groups into which a society is ...
, or tail, on one end, and a
hydrophilic A hydrophile is a molecule or other molecular entity that is attracted to water molecules and tends to be dissolved by water.Liddell, H.G. & Scott, R. (1940). ''A Greek-English Lexicon'' Oxford: Clarendon Press. In contrast, hydrophobes are no ...
moiety, or head group, on the other. The polar head group may be
anion An ion () is an atom or molecule with a net electrical charge. The charge of an electron is considered to be negative by convention and this charge is equal and opposite to the charge of a proton, which is considered to be positive by conven ...
ic,
cation An ion () is an atom or molecule with a net electrical charge. The charge of an electron is considered to be negative by convention and this charge is equal and opposite to the charge of a proton, which is considered to be positive by conven ...
ic,
zwitterion In chemistry, a zwitterion ( ; ), also called an inner salt or dipolar ion, is a molecule that contains an equal number of positively- and negatively-charged functional groups. : With amino acids, for example, in solution a chemical equilibrium w ...
ic, or non-ionic. When the concentration of a surfactant in solution reaches its
critical micelle concentration In colloidal and surface chemistry, the critical micelle concentration (CMC) is defined as the concentration of surfactants above which micelles form and all additional surfactants added to the system will form micelles. The CMC is an important c ...
(CMC), it forms micelles which are aggregates of the monomers. The CMC is different for each surfactant, as is the number of monomers which make up the micelle, termed the
aggregation number An aggregation number is a description of the number of molecules present in a micelle once the critical micelle concentration (CMC) has been reached. In more detail, it has been defined as the average number of surfactant monomers in a spherical m ...
(AN). Table 1 lists some common detergents used to form micelles along with their CMC and AN where available. Many of the characteristics of micelles differ from those of bulk solvents. For example, the micelles are, by nature, spatially
heterogeneous Homogeneity and heterogeneity are concepts often used in the sciences and statistics relating to the uniformity of a substance or organism. A material or image that is homogeneous is uniform in composition or character (i.e. color, shape, size, ...
with a hydrocarbon, nearly
anhydrous A substance is anhydrous if it contains no water. Many processes in chemistry can be impeded by the presence of water; therefore, it is important that water-free reagents and techniques are used. In practice, however, it is very difficult to achie ...
core and a highly solvated, polar head group. They have a high
surface-to-volume ratio The surface-area-to-volume ratio, also called the surface-to-volume ratio and variously denoted sa/vol or SA:V, is the amount of surface area per unit volume of an object or collection of objects. SA:V is an important concept in science and engin ...
due to their small size and generally
spherical A sphere () is a geometrical object that is a three-dimensional analogue to a two-dimensional circle. A sphere is the set of points that are all at the same distance from a given point in three-dimensional space.. That given point is the ...
shape. Their surrounding environment ( pH, ionic strength, buffer ion, presence of a co-solvent, and
temperature Temperature is a physical quantity that expresses quantitatively the perceptions of hotness and coldness. Temperature is measurement, measured with a thermometer. Thermometers are calibrated in various Conversion of units of temperature, temp ...
) has an influence on their size, shape, critical micelle concentration, aggregation number and other properties.Poole, C. Journal of Chromatography A, 1998, 807, 307-310 Another important property of micelles is the Kraft point, the temperature at which the solubility of the surfactant is equal to its CMC. For HPLC applications involving micelles, it is best to choose a surfactant with a low Kraft point and CMC. A high CMC would require a high concentration of surfactant which would increase the
viscosity The viscosity of a fluid is a measure of its resistance to deformation at a given rate. For liquids, it corresponds to the informal concept of "thickness": for example, syrup has a higher viscosity than water. Viscosity quantifies the inter ...
of the mobile phase, an undesirable condition. Additionally, a Kraft point should be well below room temperature to avoid having to apply heat to the mobile phase. To avoid potential interference with absorption detectors, a surfactant should also have a small
molar absorptivity Molar may refer to: *Molar (tooth), a kind of tooth found in mammals * Molar (grape), another name for the Spanish wine grape Listan Negro * Molar (unit), a unit of concentration equal to 1 mole per litre *Molar mass *Molar volume * El Molar, Tarra ...
at the chosen
wavelength In physics, the wavelength is the spatial period of a periodic wave—the distance over which the wave's shape repeats. It is the distance between consecutive corresponding points of the same phase on the wave, such as two adjacent crests, tro ...
of analysis.
Light scattering Scattering is a term used in physics to describe a wide range of physical processes where moving particles or radiation of some form, such as light or sound, are forced to deviate from a straight trajectory by localized non-uniformities (including ...
should not be a concern due to the small size, a few
nanometer 330px, Different lengths as in respect to the molecular scale. The nanometre (international spelling as used by the International Bureau of Weights and Measures; SI symbol: nm) or nanometer (American and British English spelling differences#-re, ...
s, of the micelle. The effect of organic additives on micellar properties is another important consideration. A small amount of organic solvent is often added to the mobile phase to help improve efficiency and to improve separations of compounds. Care needs to be taken when determining how much organic to add. Too high a concentration of the organic may cause the micelle to disperse, as it relies on hydrophobic effects for its formation. The maximum concentration of organic depends on the organic solvent itself, and on the micelle. This information is generally not known precisely, but a generally accepted practice is to keep the
volume percent In chemistry and fluid mechanics, the volume fraction φ''i'' is defined as the volume of a constituent ''V'i'' divided by the volume of all constituents of the mixture ''V'' prior to mixing: :\phi_i = \frac Being dimensionless, its unit is ...
age of organic below 15-20%.


Research

Fischer and Jandera studied the effect of changing the concentration of methanol on CMC values for three commonly used surfactants. Two cationic, hexadecyltrimethylammonium bromide (CTAB), and N-(a-carbethoxypentadecyl) trimethylammonium bromide ( Septonex), and one anionic surfactant, sodium dodecyl sulphate (SDS) were chosen for the experiment. Generally speaking, the CMC increased as the concentration of methanol increased. It was then concluded that the distribution of the surfactant between the bulk mobile phase and the micellar phase shifts toward the bulk as the methanol concentration increases. For CTAB, the rise in CMC is greatest from 0–10% methanol, and is nearly constant from 10–20%. Above 20% methanol, the micelles disaggregate and do not exist. For SDS, the CMC values remain unaffected below 10% methanol, but begin to increase as the methanol concentration is further increased. Disaggregation occurs above 30% methanol. Finally, for Septonex, only a slight increase in CMC is observed up to 20%, with disaggregation occurring above 25%. As has been asserted, the mobile phase in MLC consists of micelles in an
aqueous An aqueous solution is a solution in which the solvent is water. It is mostly shown in chemical equations by appending (aq) to the relevant chemical formula. For example, a solution of table salt, or sodium chloride (NaCl), in water would be ...
solvent, usually with a small amount of organic modifier added to complete the mobile phase. A typical reverse phase
alkyl In organic chemistry, an alkyl group is an alkane missing one hydrogen. The term ''alkyl'' is intentionally unspecific to include many possible substitutions. An acyclic alkyl has the general formula of . A cycloalkyl is derived from a cycloal ...
-bonded stationary phase is used. The first discussion of the
thermodynamics Thermodynamics is a branch of physics that deals with heat, work, and temperature, and their relation to energy, entropy, and the physical properties of matter and radiation. The behavior of these quantities is governed by the four laws of ther ...
involved in the retention mechanism was published by Armstrong and Nome in 1981. In MLC, there are three
partition coefficient In the physical sciences, a partition coefficient (''P'') or distribution coefficient (''D'') is the ratio of concentrations of a compound in a mixture of two immiscible solvents at equilibrium. This ratio is therefore a comparison of the solu ...
s which must be taken into account. The solute will partition between the water and the stationary phase (KSW), the water and the micelles (KMW), and the micelles and the stationary phase (KSM). Armstrong and Nome derived an equation describing the partition coefficients in terms of the
retention factor In chromatography, the retardation factor (''R'') is the fraction of an analyte in the mobile phase of a chromatographic system. In planar chromatography in particular, the retardation factor ''R''F is defined as the ratio of the distance traveled ...
, formally capacity factor, k¢. In HPLC, the capacity factor represents the molar ratio of the solute in the stationary phase to the mobile phase. The capacity factor is easily measure based on retention times of the compound and any unretained compound. The equation rewritten by Guermouche et al. is presented here: :1/k¢ = • (KMW-1)/(f • KSW)• CM +1/(f • KSW) Where: *k¢ is the capacity factor of the solute *KSW is the partition coefficient of the solute between the stationary phase and the water *KMW is the partition coefficient of the solute between the micelles and the water *f is the phase volume ratio (stationary phase volume/mobile phase volume) *n is the
molar volume In chemistry and related fields, the molar volume, symbol ''V''m, or \tilde V of a substance is the ratio of the volume occupied by a substance to the amount of substance, usually given at a given temperature and pressure. It is equal to the mola ...
of the surfactant *CM is the concentration of the micelle in the mobile phase (total surfactant concentration - critical micelle concentration) A
plot Plot or Plotting may refer to: Art, media and entertainment * Plot (narrative), the story of a piece of fiction Music * ''The Plot'' (album), a 1976 album by jazz trumpeter Enrico Rava * The Plot (band), a band formed in 2003 Other * ''Plot' ...
of 1/k¢ verses CM gives a straight line in which KSW can be calculated from the intercept and KMW can be obtained from the ratio of the slope to the intercept. Finally, KSM can be obtained from the ratio of the other two partition coefficients: :KSM = KSW/ KMW As can be observed from Figure 1, KMW is independent of any effects from the stationary phase, assuming the same micellar mobile phase. The validity of the retention mechanism proposed by Armstrong and Nome has been successfully, and repeated confirmed experimentally. However, some variations and alternate theories have also been proposed. Jandera and Fischer developed equations to describe the dependence of retention behavior on the change in micellar concentrations. They found that the retention of most compounds tested decreased with increasing concentrations of micelles. From this, it can be surmised that the compounds associate with the micelles as they spend less time associated with the stationary phase. Foley proposed a similar retentive
model A model is an informative representation of an object, person or system. The term originally denoted the plans of a building in late 16th-century English, and derived via French and Italian ultimately from Latin ''modulus'', a measure. Models c ...
to that of Armstrong and Nome which was a general model for secondary chemical equilibria in liquid chromatography. While this model was developed in a previous reference, and could be used for any secondary chemical equilibria such as acid-base equilibria, and ion-pairing, Foley further refined the model for MLC. When an equilibrant (X), in this case surfactant, is added to the mobile phase, a secondary equilibria is created in which an analyte will exist as free analyte (A), and complexed with the equilibrant (AX). The two forms will be retained by the stationary phase to different extents, thus allowing the retention to be varied by adjusting the concentration of equilibrant (micelles). The resulting equation solved for capacity factor in terms of partition coefficients is much the same as that of Armstrong and Nome: :1/k¢ = (KSM/k¢S) • + 1/k¢S Where: *k¢ is the capacity factor of the complexed solute and the free solute *k¢S is the capacity factor of the free solute *KSM is the partition coefficient of the solute between the stationary phase and the micelle * may be either the concentration of surfactant or the concentration of micelle Foley used the above equation to determine the solute-micelle association constants and free solute retention factors for a variety of solutes with different surfactants and stationary phases. From this data, it is possible to predict the type and optimum surfactant concentrations needed for a given solute or solutes.Foley, J.P. Analytica Chimica Acta, 1990, 231, 237-247 Foley has not been the only researcher interested in determining the solute-micelle association constants. A review article by Marina and Garcia with 53 references discusses the usefulness of obtaining solute-micelle association constants. The association constants for two solutes can be used to help understand the retention mechanism. The separation factor of two solutes, a, can be expressed as KSM1/KSM2. If the experimental a coincides with the ratio of the two solute-micelle partition coefficients, it can be assumed that their retention occurs through a direct transfer from the micellar phase to the stationary phase. In addition, calculation of a would allow for prediction of separation selectivity before the analysis is performed, provided the two coefficients are known. The desire to predict retention behavior and selectivity has led to the development of several mathematical models. Changes in pH, surfactant concentration, and concentration of organic modifier play a significant role in determining the chromatographic separation. Often one or more of these parameters need to be optimized to achieve the desired separation, yet the optimum parameters must take all three variables into account simultaneously. The review by Garcia-Alvarez-Coque et al. mentioned several successful models for varying scenarios, a few of which will be mentioned here. The classic models by Armstrong and Nome and Foley are used to describe the general cases. Foley's model applies to many cases and has been experimentally verified for ionic, neutral, polar and nonpolar solutes; anionic, cationic, and non-ionic surfactants, and C8, C¬18, and
cyano Cyanide is a naturally occurring, rapidly acting, toxic chemical that can exist in many different forms. In chemistry, a cyanide () is a chemical compound that contains a functional group. This group, known as the cyano group, consists of a ...
stationary phases. The model begins to deviate for highly and lowly retained solutes. Highly retained solutes may become irreversibly bound to the stationary phase, where lowly retained solutes may elute in the column void volume. Other models proposed by Arunyanart and Cline-Love and Rodgers and Khaledi describe the effect of pH on the retention of weak acids and bases. These authors derived equations relating pH and micellar concentration to retention. As the pH varies, sigmoidal behavior is observed for the retention of acidic and basic species. This model has been shown to accurately predict retention behavior. Still other models predict behavior in hybrid micellar systems using equations or modeling behavior based on controlled experimentation. Additionally, models accounting for the simultaneous effect of pH, micelle and organic concentration have been suggested. These models allow for further enhancement of the optimization of the separation of weak acids and bases.Garcia-Alvarez-Coque, M.C.; Torres-Lapasio, J.R.; Baeza-Baeza, J.J.; Journal of Chromatography A, 1997, 780, 129-148 One research group, Rukhadze, et al. derived a first order linear relationship describing the influence of micelle and organic concentration, and pH on the selectivity and resolution of seven
barbiturate Barbiturates are a class of depressant drugs that are chemically derived from barbituric acid. They are effective when used medically as anxiolytics, hypnotics, and anticonvulsants, but have physical and psychological addiction potential as we ...
s. The researchers discovered that a second order mathematical equation would more precisely fit the data. The derivations and experimental details are beyond the scope of this discussion. The model was successful in predicting the experimental conditions necessary to achieve a separation for compounds which are traditionally difficult to resolve. Jandera, Fischer, and Effenberger approached the modeling problem in yet another way. The model used was based on
lipophilicity Lipophilicity (from Greek λίπος "fat" and φίλος "friendly"), refers to the ability of a chemical compound to dissolve in fats, oils, lipids, and non-polar solvents such as hexane or toluene. Such non-polar solvents are themselves lipo ...
and polarity indices of solutes. The lipophilicity index relates a given solute to a hypothetical number of carbon atoms in an alkyl chain. It is based and depends on a given calibration series determined experimentally. The lipophilicity index should be independent of the stationary phase and organic modifier concentration. The polarity index is a measure of the polarity of the solute-solvent interactions. It depends strongly on the organic solvent, and somewhat on the polar groups present in the stationary phase. 23 compounds were analyzed with varying mobile phases and compared to the lipophilicity and polarity indices. The results showed that the model could be applied to MLC, but better predictive behavior was found with concentrations of surfactant below the CMC, sub-micellar. A final type of model based on molecular properties of a solute is a branch of
quantitative structure-activity relationship Quantitative may refer to: * Quantitative research, scientific investigation of quantitative properties * Quantitative analysis (disambiguation) * Quantitative verse, a metrical system in poetry * Statistics, also known as quantitative analysis ...
s (QSAR). QSAR studies attempt to correlate biological activity of
drug A drug is any chemical substance that causes a change in an organism's physiology or psychology when consumed. Drugs are typically distinguished from food and substances that provide nutritional support. Consumption of drugs can be via inhalati ...
s, or a class of drugs, with structures. The normally accepted means of uptake for a drug, or its metabolite, is through partitioning into
lipid bilayer The lipid bilayer (or phospholipid bilayer) is a thin polar membrane made of two layers of lipid molecules. These membranes are flat sheets that form a continuous barrier around all cells. The cell membranes of almost all organisms and many vi ...
s. The descriptor most often used in QSAR to determine the hydrophobicity of a compound is the
octanol Octanols are alcohols with the formula C8H17OH. A simple and important member is 1-octanol, with an unbranched chain of carbons. Other commercially important octanols are 2-octanol and 2-ethylhexanol. There are 89 possible isomers In che ...
-water partition coefficient,
log Log most often refers to: * Trunk (botany), the stem and main wooden axis of a tree, called logs when cut ** Logging, cutting down trees for logs ** Firewood, logs used for fuel ** Lumber or timber, converted from wood logs * Logarithm, in mat ...
P. MLC provides an attractive and practical alternative to QSAR. When micelles are added to a mobile phase, many similarities exist between the micellar mobile phase/stationary phase and the biological membrane/water interface. In MLC, the stationary phase become modified by the adsorption of surfactant monomers which are structurally similar to the membranous hydrocarbon chains in the biological model. Additionally, the hydrophilic/hydrophobic interactions of the micelles are similar to that in the polar regions of a membrane. Thus, the development of quantitative structure-retention relationships (QRAR) has become widespread. Escuder-Gilabert et al. tested three different QRAR retention models on ionic compounds. Several classes of compounds were tested including
catecholamine A catecholamine (; abbreviated CA) is a monoamine neurotransmitter, an organic compound that has a catechol (benzene with two hydroxyl side groups next to each other) and a side-chain amine. Catechol can be either a free molecule or a substi ...
s,
local anesthetic A local anesthetic (LA) is a medication that causes absence of pain sensation. In the context of surgery, a local anesthetic creates an absence of pain in a specific location of the body without a loss of consciousness, as opposed to a general a ...
s,
diuretic A diuretic () is any substance that promotes diuresis, the increased production of urine. This includes forced diuresis. A diuretic tablet is sometimes colloquially called a water tablet. There are several categories of diuretics. All diuretics in ...
s, and
amino acid Amino acids are organic compounds that contain both amino and carboxylic acid functional groups. Although hundreds of amino acids exist in nature, by far the most important are the alpha-amino acids, which comprise proteins. Only 22 alpha am ...
s. The best model relating log K and log P was found to be one in which the total molar charge of a compound at a given pH is included as a variable. This model proved to give fairly accurate predictions of log P, R > 0.9. Other studies have been performed which develop predictive QRAR models for tricyclic antidepressants and barbiturates.Cuenca-Benito, M.; Sagrado, S.; Villanueva-Camanas, R.; Medina-Hernandez, M; Journal of Chromatography A, 1998, 814, 121-132


Efficiency

The main limitation in the use of MLC is the reduction in efficiency (peak broadening) that is observed when purely aqueous micellar mobile phases are used. Several explanations for the poor efficiency have been theorized. Poor wetting of the stationary phase by the micellar aqueous mobile phase, slow
mass transfer Mass transfer is the net movement of mass from one location (usually meaning stream, phase, fraction or component) to another. Mass transfer occurs in many processes, such as absorption, evaporation, drying, precipitation, membrane filtration ...
between the micelles and the stationary phase, and poor mass transfer within the stationary phase have all been postulated as possible causes. To enhance efficiency, the most common approaches have been the addition of small amounts of
isopropyl alcohol Isopropyl alcohol (IUPAC name propan-2-ol and also called isopropanol or 2-propanol) is a colorless, flammable organic compound with a pungent alcoholic odor. As an isopropyl group linked to a hydroxyl group ( chemical formula ) it is the simpl ...
and increase in temperature. A review by Berthod studied the combined theories presented above and applied the Knox equation to independently determine the cause of the reduced efficiency. The Knox equation is commonly used in HPLC to describe the different contributions to overall band broadening of a solute. The Knox equation is expressed as: :h = An^(1/3)+ B/n + Cn Where: *h = the reduced plate height count (plate height/stationary phase particle diameter) *n = the reduced mobile phase linear velocity (velocity times stationary phase particle diameter/solute diffusion coefficient in the mobile phase) *A, B, and C are constants related to solute flow anisotropy (eddy diffusion), molecular longitudinal diffusion, and mass transfer properties respectively. Berthod's use of the Knox equation to experimentally determine which of the proposed theories was most correct led him to the following conclusions. The flow
anisotropy Anisotropy () is the property of a material which allows it to change or assume different properties in different directions, as opposed to isotropy. It can be defined as a difference, when measured along different axes, in a material's physi ...
in micellar phase seems to be much greater than in traditional hydro-organic mobile phases of similar
viscosity The viscosity of a fluid is a measure of its resistance to deformation at a given rate. For liquids, it corresponds to the informal concept of "thickness": for example, syrup has a higher viscosity than water. Viscosity quantifies the inter ...
. This is likely due to the partial clogging of the stationary phase pores by adsorbed surfactant molecules. Raising the column temperature served to both decrease viscosity of the mobile phase and the amount of adsorbed surfactant. Both results reduce the A term and the amount of
eddy diffusion Eddy diffusion, eddy dispersion, or turbulent diffusion is a process by which substances are mixed in the atmosphere, the ocean or in any fluid system due to eddy motion. In other words, it is mixing that is caused by eddies that can vary in size f ...
, and thereby increase efficiency. The increase in the B term, as related to longitudinal diffusion, is associated with the decrease in the solute diffusion coefficient in the mobile phase, DM, due to the presence of the micelles, and an increase in the capacity factor, k¢. Again, this is related to surfactant adsorption on the stationary phase causing a dramatic decrease in the solute diffusion coefficient in the stationary phase, DS. Again an increase in temperature, now coupled with an addition of alcohol to the mobile phase, drastically decreases the amount of the absorbed surfactant. In turn, both actions reduce the C term caused by a slow mass transfer from the stationary phase to the mobile phase. Further optimization of efficiency can be gained by reducing the flow rate to one closely matched to that derived from the Knox equation. Overall, the three proposed theories seemed to have contributing effects of the poor efficiency observed, and can be partially countered by the addition of organic modifiers, particularly alcohol, and increasing the column temperature.


Applications

Despite the reduced efficiency verses reversed phase HPLC, hundreds of applications have been reported using MLC. One of the most advantageous is the ability to directly inject physiological fluids. Micelles have an ability to solubilize proteins which enables MLC to be useful in analyzing untreated biological fluids such as plasma, serum, and
urine Urine is a liquid by-product of metabolism in humans and in many other animals. Urine flows from the kidneys through the ureters to the urinary bladder. Urination results in urine being excreted from the body through the urethra. Cellular m ...
. Martinez et al. found MLC to be highly useful in analyzing a class of drugs called b-antagonists, so called
beta-blockers Beta blockers, also spelled β-blockers, are a class of medications that are predominantly used to manage abnormal heart rhythms, and to protect the heart from a second heart attack after a first heart attack (secondary prevention). They are al ...
, in urine samples. The main advantage of the use of MLC with this type of sample, is the great time savings in sample preparation. Alternative methods of analysis including reversed phase HPLC require lengthy extraction and sample work up procedures before analysis can begin. With MLC, direct injection is often possible, with retention times of less than 15 minutes for the separation of up to nine b-antagonists. Another application compared reversed phase HPLC with MLC for the analysis of desferrioxamine in serum. Desferrioxamine (DFO) is a commonly used drug for removal of excess iron in patients with chronic and acute levels. The analysis of DFO along with its chelated complexes, Fe(III) DFO and Al(III) DFO has proven to be difficult at best in previous attempts. This study found that direct injection of the serum was possible for MLC, verses an
ultrafiltration Ultrafiltration (UF) is a variety of membrane filtration in which forces such as pressure or concentration gradients lead to a separation through a semipermeable membrane. Suspended solids and solutes of high molecular weight are retained in the ...
step necessary in HPLC. This analysis proved to have difficulties with the separation of the chelated DFO compounds and with the sensitivity levels for DFO itself when MLC was applied. The researcher found that, in this case, reverse phase HPLC, was a better, more sensitive technique despite the time savings in direct injection. Analysis of pharmaceuticals by MLC is also gaining popularity. The selectivity and peak shape of MLC over commonly used ion-pair chromatography is much enhanced. MLC mimics, yet enhances, the selectivity offered by ion-pairing reagents for the separation of active ingredients in
pharmaceutical drug A medication (also called medicament, medicine, pharmaceutical drug, medicinal drug or simply drug) is a drug used to diagnose, cure, treat, or prevent disease. Drug therapy (pharmacotherapy) is an important part of the medical field and ...
s. For basic drugs, MLC improves the excessive peak tailing frequently observed in ion-pairing.
Hydrophilic A hydrophile is a molecule or other molecular entity that is attracted to water molecules and tends to be dissolved by water.Liddell, H.G. & Scott, R. (1940). ''A Greek-English Lexicon'' Oxford: Clarendon Press. In contrast, hydrophobes are no ...
drugs are often unretained using conventional HPLC, are retained by MLC due to solubilization into the micelles. Commonly found drugs in cold medications such as
acetaminophen Paracetamol, also known as acetaminophen, is a medication used to treat fever and mild to moderate pain. Common brand names include Tylenol and Panadol. At a standard dose, paracetamol only slightly decreases body temperature; it is inferior ...
, L-ascorbic acid,
phenylpropanolamine Phenylpropanolamine (PPA) is a sympathomimetic agent which is used as a decongestant and appetite suppressant. It was commonly used in prescription and over-the-counter cough and cold preparations. In veterinary medicine, it is used to contro ...
HCL, tipepidine hibenzate, and chlorpheniramine maleate have been successfully separated with good peak shape using MLC. Additional basic drugs like many narcotics, such as
codeine Codeine is an opiate and prodrug of morphine mainly used to treat pain, coughing, and diarrhea. It is also commonly used as a recreational drug. It is found naturally in the sap of the opium poppy, ''Papaver somniferum''. It is typically used ...
and
morphine Morphine is a strong opiate that is found naturally in opium, a dark brown resin in poppies (''Papaver somniferum''). It is mainly used as a pain medication, and is also commonly used recreationally, or to make other illicit opioids. There ...
, have also been successfully separated using MLC. Another novel application of MLC involves the separation and analysis of
inorganic compound In chemistry, an inorganic compound is typically a chemical compound that lacks carbon–hydrogen bonds, that is, a compound that is not an organic compound. The study of inorganic compounds is a subfield of chemistry known as ''inorganic chemistr ...
s, mostly simple ions. This is a relatively new area for MLC, but has seen some promising results. MLC has been observed to provide better selectivity of inorganic ions that ion-exchange or ion-pairing chromatography. While this application is still in the beginning stages of development, the possibilities exist for novel, much enhanced separations of inorganic species. Since the technique was first reported on in 1980, micellar liquid chromatography has been used in hundreds of applications. This micelle controlled technique provides for unique opportunities for solving complicated separation problems. Despite the poor efficiency of MLC, it has been successfully used in many applications. The use of MLC in the future appears to be extremely advantages in the areas of physiological fluids, pharmaceuticals, and even inorganic ions. The technique has proven to be superior over ion-pairing and ion-exchange for many applications. As new approaches are developed to combat the poor efficiency of MLC, its application is sure to spread and gain more acceptance.


References

{{reflist, 2 Chromatography