Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or mass spectrometer) for determining the
mass-to-charge ratio The mass-to-charge ratio (''m''/''Q'') is a physical quantity relating the ''mass'' (quantity of matter) and the ''electric charge'' of a given particle, expressed in units of kilograms per coulomb (kg/C). It is most widely used in the electrody ...
(''m''/''z'') of ions based on the cyclotron frequency of the ions in a fixed magnetic field. The ions are trapped in a Penning trap (a magnetic field with electric trapping plates), where they are excited (at their resonant cyclotron frequencies) to a larger cyclotron radius by an oscillating electric field orthogonal to the magnetic field. After the excitation field is removed, the ions are rotating at their cyclotron frequency in phase (as a "packet" of ions). These ions induce a charge (detected as an image current) on a pair of electrodes as the packets of ions pass close to them. The resulting signal is called a free induction decay (FID), transient or interferogram that consists of a superposition of sine waves. The useful signal is extracted from this data by performing a
Fourier transform A Fourier transform (FT) is a mathematical transform that decomposes functions into frequency components, which are represented by the output of the transform as a function of frequency. Most commonly functions of time or space are transformed, ...
to give a mass spectrum.


History

FT-ICR was invented by Melvin B. Comisarow and Alan G. Marshall at the University of British Columbia. The first paper appeared in Chemical Physics Letters in 1974. The inspiration was earlier developments in conventional ICR and Fourier-transform nuclear magnetic resonance (FT-NMR) spectrometry. Marshall has continued to develop the technique at The Ohio State University and
Florida State University Florida State University (FSU) is a public research university in Tallahassee, Florida. It is a senior member of the State University System of Florida. Founded in 1851, it is located on the oldest continuous site of higher education in the st ...
.


Theory

The physics of FTICR is similar to that of a cyclotron at least in the first approximation. In the simplest idealized form, the relationship between the cyclotron frequency and the mass-to-charge ratio is given by :f = \frac, where ''f'' = cyclotron frequency, ''q'' = ion charge, ''B'' = magnetic field strength and ''m'' = ion mass. This is more often represented in angular frequency: :\omega_\text = \frac, where \omega_\text is the angular cyclotron frequency, which is related to frequency by the definition f = \frac. Because of the quadrupolar electrical field used to trap the ions in the axial direction, this relationship is only approximate. The axial electrical trapping results in axial oscillations within the trap with the (angular) frequency :\omega_\text = \sqrt, where \alpha is a constant similar to the spring constant of a
harmonic oscillator In classical mechanics, a harmonic oscillator is a system that, when displaced from its Mechanical equilibrium, equilibrium position, experiences a restoring force ''F'' Proportionality (mathematics), proportional to the displacement ''x'': \v ...
and is dependent on applied voltage, trap dimensions and trap geometry. The electric field and the resulting axial harmonic motion reduces the cyclotron frequency and introduces a second radial motion called magnetron motion that occurs at the magnetron frequency. The cyclotron motion is still the frequency being used, but the relationship above is not exact due to this phenomenon. The natural angular frequencies of motion are :\omega_\pm = \frac \pm \sqrt, where \omega_\text is the axial trapping frequency due the axial electrical trapping and \omega_+ is the reduced cyclotron (angular) frequency and \omega_- is the magnetron (angular) frequency. Again, \omega_+ is what is typically measured in FTICR. The meaning of this equation can be understood qualitatively by considering the case where \omega_\text is small, which is generally true. In that case the value of the radical is just slightly less than \omega_\text/2, and the value of \omega_+ is just slightly less than \omega_\text (the cyclotron frequency has been slightly reduced). For \omega_- the value of the radical is the same (slightly less than \omega_\text/2), but it is being subtracted from \omega_\text/2, resulting in a small number equal to \omega_\text - \omega_+ (i.e. the amount that the cyclotron frequency was reduced by).


Instrumentation

FTICR-MS differs significantly from other
mass spectrometry Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a ''mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is use ...
techniques in that the ions are not detected by hitting a detector such as an electron multiplier but only by passing near detection plates. Additionally the masses are not resolved in space or time as with other techniques but only by the ion cyclotron resonance (rotational) frequency that each ion produces as it rotates in a magnetic field. Thus, the different ions are not detected in different places as with
sector instrument A sector instrument is a general term for a class of mass spectrometer that uses a static electric (E) or magnetic (B) sector or some combination of the two (separately in space) as a mass analyzer. Popular combinations of these sectors have been ...
s or at different times as with time-of-flight instruments, but all ions are detected simultaneously during the detection interval. This provides an increase in the observed
signal-to-noise ratio Signal-to-noise ratio (SNR or S/N) is a measure used in science and engineering that compares the level of a desired signal to the level of background noise. SNR is defined as the ratio of signal power to the noise power, often expressed in deci ...
owing to the principles of
Fellgett's advantage Fellgett's advantage or the multiplex advantage is an improvement in signal-to-noise ratio (SNR) that is gained when taking multiplexed measurements rather than direct measurements. The name is derived from P. B. Fellgett, who first made the observ ...
. In FTICR-MS, resolution can be improved either by increasing the strength of the magnet (in teslas) or by increasing the detection duration.


Cells

A review of different cell geometries with their specific electric configurations is available in the literature. However, ICR cells can belong to one of the following two categories: closed cells or open cells. Several closed ICR cells with different geometries were fabricated and their performance has been characterized. Grids were used as end caps to apply an axial electric field for trapping ions axially (parallel to the magnetic field lines). Ions can be either generated inside the cell or can be injected to the cell from an external ionization source. Nested ICR cells with double pair of grids were also fabricated to trap both positive and negative ions simultaneously. The most common open cell geometry is a cylinder, which is axially segmented to produce electrodes in the shape of a ring. The central ring electrode is commonly used for applying radial excitation electric field and detection. DC electric voltage is applied on the terminal ring electrodes to trap ions along the magnetic field lines. Open cylindrical cells with ring electrodes of different diameters have also been designed. They proved not only capable in trapping and detecting both ion polarities simultaneously, but also they succeeded to separate positive from negative ions radially. This presented a large discrimination in kinetic ion acceleration between positive and negative ions trapped simultaneously inside the new cell. Several ion axial acceleration schemes were recently written for ion–ion collision studies.


Stored-waveform inverse Fourier transform

Stored-waveform inverse Fourier transform (SWIFT) is a method for the creation of excitation waveforms for FTMS. The time-domain excitation waveform is formed from the inverse Fourier transform of the appropriate frequency-domain excitation spectrum, which is chosen to excite the resonance frequencies of selected ions. The SWIFT procedure can be used to select ions for tandem mass spectrometry experiments.


Applications

Fourier-transform ion cyclotron resonance (FTICR) mass spectrometry is a high-resolution technique that can be used to determine masses with high accuracy. Many applications of FTICR-MS use this mass accuracy to help determine the composition of molecules based on accurate mass. This is possible due to the mass defect of the elements. FTICR-MS is able to achieve higher levels of mass accuracy than other forms of mass spectrometer, in part, because a superconducting magnet is much more stable than radio-frequency (RF) voltage. Another place that FTICR-MS is useful is in dealing with complex mixtures, such as biomass or waste liquefaction products, since the resolution (narrow peak width) allows the signals of two ions with similar mass-to-charge ratios (''m''/''z'') to be detected as distinct ions. This high resolution is also useful in studying large macromolecules such as proteins with multiple charges, which can be produced by electrospray ionization. For example, attomole level of detection of two peptides has been reported. These large molecules contain a distribution of isotopes that produce a series of isotopic peaks. Because the isotopic peaks are close to each other on the ''m''/''z'' axis, due to the multiple charges, the high resolving power of the FTICR is extremely useful. FTICR-MS is very useful in other studies of proteomics as well. It achieves exceptional resolution in both top-down and bottom-up proteomics. Electron-capture dissociation (ECD), collisional-induced dissociation (CID), and infrared multiphoton dissociation (IRMPD) are all utilized to produce fragment spectra in tandem mass spectrometry experiments. Although CID and IRMPD use vibrational excitation to further dissociate peptides by breaking the backbone amide linkages, which are typically low in energy and weak, CID and IRMPD may also cause dissociation of post-translational modifications. ECD, on the other hand, allows specific modifications to be preserved. This is quite useful in analyzing phosphorylation states, O- or N-linked glycosylation, and sulfating.


References


External links


What's in an Oil Drop? An Introduction to Fourier Transform Ion Cyclotron Resonance (FT-ICR) for Non-scientists
National High Magnetic Field Laboratory
Scottish Instrumentation Resource Centre for Advanced Mass SpectrometryFourier-transform Ion Cyclotron Resonance (FT-ICR)
FT-ICR Introduction University of Bristol {{DEFAULTSORT:Fourier Transform Ion Cyclotron Resonance Mass spectrometry Measuring instruments