There are a variety of technologies available for detecting and recording the images, diffraction patterns, and electron energy loss spectra produced using
transmission electron microscopy Transmission electron microscopy (TEM) is a microscopy technique in which a beam of electrons is transmitted through a specimen to form an image. The specimen is most often an ultrathin section less than 100 nm thick or a suspension on a ...
(TEM).


Traditional detection techniques

Traditionally,TEM image or diffraction pattern could be observed using a fluorescent viewing screen, consisting of powdered ZnS or ZnS/CdS, which is excited by the electron beam via
cathodoluminescence Cathodoluminescence is an optical and electromagnetic phenomenon in which electrons impacting on a luminescent material such as a phosphor, cause the emission of photons which may have wavelengths in the visible spectrum. A familiar example is ...
. Once the microscopist could see a suitable image on their viewing screen, images could then be recorded using
photographic film Photographic film is a strip or sheet of transparent film base coated on one side with a gelatin emulsion containing microscopically small light-sensitive silver halide crystals. The sizes and other characteristics of the crystals determine ...
. For electron microscopes, film typically consisted of a gelatin and silver halide emulsion layer on a plastic support base. The silver halide would be converted to silver upon exposure to the electron beam, and the film could then be chemically developed to form an image, which could be digitized for analysis using a film scanner. In modern TEMs, film has largely been replaced by electronic detectors.


CCD cameras

Charge coupled device A charge-coupled device (CCD) is an integrated circuit containing an array of linked, or coupled, capacitors. Under the control of an external circuit, each capacitor can transfer its electric charge to a neighboring capacitor. CCD sensors are a ...
(CCD) cameras were first applied to transmission electron microscopy in the 1980s and later became widespread. For use in a TEM, CCDs are typically coupled with a
scintillator A scintillator is a material that exhibits scintillation, the property of luminescence, when excited by ionizing radiation. Luminescent materials, when struck by an incoming particle, absorb its energy and scintillate (i.e. re-emit the absorbe ...
such as single crystal
Yttrium aluminium garnet Yttrium aluminium garnet (YAG, Y3 Al5 O12) is a synthetic crystalline material of the garnet group. It is a cubic yttrium aluminium oxide phase, with other examples being YAlO3 (YAP) in a hexagonal or an orthorhombic, perovskite-like form, and ...
(YAG) in which electrons from the electron beam are converted to photons, which are then transferred to the sensor of the CCD via a fiber optic plate. The main reason for this is that direct exposure to the high energy electron beam risks damaging the sensor CCD. A typical CCD for a TEM will also incorporate a Peltier cooling device to reduce the temperature of the sensor to approximately -30 °C, which reduces dark current and improves signal to noise.


CMOS cameras

Since 2006, scintillator and fiber optic coupled cameras based on
complementary metal oxide semiconductor Complementary metal–oxide–semiconductor (CMOS, pronounced "sea-moss", ) is a type of metal–oxide–semiconductor field-effect transistor (MOSFET) fabrication process that uses complementary and symmetrical pairs of p-type and n-type MOSFE ...
(CMOS) electronics have become commercially available for TEM. CMOS cameras have some advantages for electron microscopy compared to CCD cameras. One advantage is that CMOS cameras are less prone than CCD cameras to blooming, i.e. the spreading of charge from oversaturated pixels into nearby pixels. Another advantage is that CMOS cameras can have faster readout speeds.


Direct electron detectors

The use of scintillators to convert electrons to photons in CCD and CMOS cameras reduces the detective quantum efficiency (DQE) of these devices. Direct electron detectors, which have no scintillator and are directly exposed to the electron beam, typically offer higher DQE than scintillator coupled cameras. There are two main types of direct electron detector, both of which were first introduced to electron microscopy in the 2000s. A
hybrid pixel detector Hybrid pixel detectors are a type of ionizing radiation detector consisting of an array of diodes based on semiconductor technology and their associated electronics. The term “hybrid” stems from the fact that the two main elements from which th ...
, also known as a pixel array detector (PAD) features a sensor chip bonded to a separate electronics chip with each pixel read out in parallel. The pixels are typically wide and thick e.g. 150 x 150 x 500 µm for the electron microscope pixel array detector (EMPAD) described by Tate et al. This large pixel size allows each pixel to fully absorb high energy electrons, enabling high dynamic range. However, the large pixel size limits the number of pixels that can be incorporated into a sensor. A monolithic
active pixel sensor An active-pixel sensor (APS) is an image sensor where each pixel sensor unit cell has a photodetector (typically a pinned photodiode) and one or more active transistors. In a metal–oxide–semiconductor (MOS) active-pixel sensor, MOS field-effec ...
(MAPS) for TEM is a CMOS based detector that has been radiation hardened to withstand direct exposure to the electron beam. The sensitive layer of the MAPS is typically very thin, with a thickness as low as 8 μm. This reduces the lateral spread of electrons from the electron beam within the detective layer of the sensor, allowing for smaller pixel sizes e.g. 6.5 x 6.5 µm for a Direct Electron DE-16. Smaller pixel size allows for a large number of pixels to be incorporated into a sensor, although dynamic range is typically more limited than for a hybrid pixel detector.


Detectors for Scanning TEM (STEM)

In scanning TEM (STEM), a focused probe is rastered over an area of interest, and a signal is recorded at each probe position to form an image. This typically requires different types of detector from conventional TEM imaging, in which a broad area of the specimen is illuminated. Traditional STEM imaging involves detectors, such as the annular dark-field (ADF) detector, which integrate the signal resulting from electrons from within a given range of scattering angles at each position of the raster. Such detectors may typically consist of a scintillator connected to
photomultiplier A photomultiplier is a device that converts incident photons into an electrical signal. Kinds of photomultiplier include: * Photomultiplier tube, a vacuum tube converting incident photons into an electric signal. Photomultiplier tubes (PMTs for sh ...
tube. Segmented STEM detectors, first introduced in 1994 allow differential phase contrast information to be obtained. 4D STEM involves the use of an imaging camera, such as they hybrid pixel or MAPS direct electron detectors described above, to record an entire
convergent beam electron diffraction Convergent beam electron diffraction (CBED) is a diffraction technique where a convergent or divergent beam (conical electron beam) of electrons is used to study materials. History This technique was first introduced in 1939 by Kossel and Möll ...
(CBED) pattern at each STEM raster position. The resulting four-dimensional dataset can then be analyzed to reconstruct arbitrary STEM images, or extract other types of information from the specimen, such as strain, or electric and magnetic field maps.