Insulation monitoring device
   HOME

TheInfoList



OR:

An insulation monitoring device monitors the ungrounded system between an
active phase conductor Active may refer to: Music * ''Active'' (album), a 1992 album by Casiopea * Active Records, a record label Ships * ''Active'' (ship), several commercial ships by that name * HMS ''Active'', the name of various ships of the British Royal ...
and
earth Earth is the third planet from the Sun and the only astronomical object known to harbor life. While large volumes of water can be found throughout the Solar System, only Earth sustains liquid surface water. About 71% of Earth's sur ...
. It is intended to give an alert (light and sound) or disconnect the power supply when the resistance between the two conductors drops below a set value, usually 50 k Ω (sample of IEC standard for medical applications). The main advantage is that the ungrounded or floating system allows a continuous operation of important consumers such as medical, chemical, military, etc. Some manufacturers of monitors for these systems are capable of handling VFDs ( Variable Frequency (Speed) Drives). Most, however are not due to issues with the DC-portions of the VFDs. Most monitors work by injecting low level DC on the line and detecting. Some manufacturers use a patented AMP-monitoring principle (Adapted Measuring Pulse)


Insulation Monitoring Devices (IMD) according to IEC 61557-8

This part of IEC 61557 specifies requirements for insulation monitoring devices (IMD), which permanently monitor the insulation resistance ''R''F to earth of unearthed a.c. IT systems, of a.c. IT systems with galvanically connected d.c. circuits having nominal voltages up to 1 000 V a.c., as well as of unearthed d.c. IT systems with voltages up to 1 500 V d.c. independent from the method of measuring. According to 61557-8: Insulation monitoring devices must be able to monitor both symmetrical and asymmetrical insulation faults according to their specified measuring principle. A symmetrical fault in insulation occurs when the insulation resistance of all conductors in the network to be monitored is (approximately) evenly reduced. An unsymmetrical fault in insulation occurs when the insulation resistance, e.g. of one conductor is reduced much more than that of the remaining conductor(s). Relays, which use the unbalance voltage (3 voltmeter method), are not insulation monitoring devices according to this standard. Under special network conditions, a combination of several measurement methods may be necessary to fulfill the monitoring task, including an asymmetry monitoring.


Terminology

Terminology differs by region and by market segment. * For the system type different terms are used: "IT system" (lat. Isolation Terra) vs. "floating system" vs. "ungrounded system". * For the monitoring device different terms are used: "Insulation monitoring device" vs. "ground fault relay" vs. "ground fault detector" vs. "LIM" (Line Isolation Monitor)


Standards

*
IEC The International Electrotechnical Commission (IEC; in French: ''Commission électrotechnique internationale'') is an international standards organization that prepares and publishes international standards for all electrical, electronic and r ...
61557-8:2014, "Electrical safety in low voltage distribution systems up to 1 000 V a.c. and 1 500 V d.c. - Equipment for testing, measuring or monitoring of protective measures - Part 8: Insulation monitoring devices for IT systems" *
IEC The International Electrotechnical Commission (IEC; in French: ''Commission électrotechnique internationale'') is an international standards organization that prepares and publishes international standards for all electrical, electronic and r ...
61557-9:2014, "Electrical safety in low voltage distribution systems up to 1000 V a.c. and 1500 V d.c. - Equipment for testing, measuring or monitoring of protective measures - Part 9: Equipment for insulation fault location in IT systems"


Literature

* Wolfgang Hofheinz: ''Protective Measures with Insulation Monitoring'' VDE ''Application of Unearthed IT Power Systems in Industry, Mining, Railways, Marine/ Oil and Electric/Rail Vehicles'', , E-Book:


Further reading


IEC 61557-8 Electrical safety in low voltage distribution systems up to 1 000 V a.c. and 1 500 V d.c. – Equipment for testing, measuring or monitoring of protective measures – Part 8: Insulation monitoring devices for IT systems

The IT earthing system (unearthed neutral) in LVUngrounded Power Supply Systems in Medical Sites


References

{{DEFAULTSORT:Insulation Monitoring Device Electric power systems components Safety switches Electrical safety