Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or
mass spectrometer Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a '' mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is us ...
) for determining the
mass-to-charge ratio The mass-to-charge ratio (''m''/''Q'') is a physical quantity Ratio, relating the ''mass'' (quantity of matter) and the ''electric charge'' of a given particle, expressed in Physical unit, units of kilograms per coulomb (kg/C). It is most widely ...
(''m''/''z'') of
ions An ion () is an atom or molecule with a net electrical charge. The charge of an electron is considered to be negative by convention and this charge is equal and opposite to the charge of a proton, which is considered to be positive by convent ...
based on the cyclotron frequency of the ions in a fixed magnetic field. The ions are trapped in a
Penning trap A Penning trap is a device for the storage of charged particles using a homogeneous magnetic field and a quadrupole electric field. It is mostly found in the physical sciences and related fields of study for precision measurements of properties o ...
(a magnetic field with electric trapping plates), where they are excited (at their resonant cyclotron frequencies) to a larger cyclotron radius by an oscillating electric field orthogonal to the magnetic field. After the excitation field is removed, the ions are rotating at their cyclotron frequency in phase (as a "packet" of ions). These ions induce a charge (detected as an image current) on a pair of electrodes as the packets of ions pass close to them. The resulting signal is called a free induction decay (FID), transient or interferogram that consists of a superposition of
sine waves A sine wave, sinusoidal wave, or sinusoid (symbol: ∿) is a periodic wave whose waveform (shape) is the trigonometric sine function. In mechanics, as a linear motion over time, this is '' simple harmonic motion''; as rotation, it correspond ...
. The useful signal is extracted from this data by performing a
Fourier transform In mathematics, the Fourier transform (FT) is an integral transform that takes a function as input then outputs another function that describes the extent to which various frequencies are present in the original function. The output of the tr ...
to give a
mass spectrum A mass spectrum is a histogram plot of intensity vs. ''mass-to-charge ratio'' (''m/z'') in a chemical sample, usually acquired using an instrument called a ''mass spectrometer''. Not all mass spectra of a given substance are the same; for example ...
.


History

FT-ICR was invented by Melvin B. Comisarow and Alan G. Marshall at the
University of British Columbia The University of British Columbia (UBC) is a Public university, public research university with campuses near University of British Columbia Vancouver, Vancouver and University of British Columbia Okanagan, Kelowna, in British Columbia, Canada ...
. The first paper appeared in Chemical Physics Letters in 1974. The inspiration was earlier developments in conventional ICR and Fourier-transform nuclear magnetic resonance (FT-NMR) spectrometry. Marshall has continued to develop the technique at
The Ohio State University The Ohio State University (Ohio State or OSU) is a public university, public Land-grant university, land-grant research university in Columbus, Ohio, United States. A member of the University System of Ohio, it was founded in 1870. It is one ...
and
Florida State University Florida State University (FSU or Florida State) is a Public university, public research university in Tallahassee, Florida, United States. It is a senior member of the State University System of Florida and a preeminent university in the s ...
.


Theory

The physics of FTICR is similar to that of a
cyclotron A cyclotron is a type of particle accelerator invented by Ernest Lawrence in 1929–1930 at the University of California, Berkeley, and patented in 1932. Lawrence, Ernest O. ''Method and apparatus for the acceleration of ions'', filed: Januar ...
at least in the first approximation. In the simplest idealized form, the relationship between the cyclotron frequency and the mass-to-charge ratio is given by :f = \frac, where ''f'' = cyclotron frequency, ''q'' = ion charge, ''B'' =
magnetic field strength A magnetic field (sometimes called B-field) is a physical field that describes the magnetic influence on moving electric charges, electric currents, and magnetic materials. A moving charge in a magnetic field experiences a force perpendicula ...
and ''m'' = ion mass. This is more often represented in
angular frequency In physics, angular frequency (symbol ''ω''), also called angular speed and angular rate, is a scalar measure of the angle rate (the angle per unit time) or the temporal rate of change of the phase argument of a sinusoidal waveform or sine ...
: :\omega_\text = \frac, where \omega_\text is the angular cyclotron frequency, which is related to frequency by the definition f = \frac. Because of the quadrupolar electrical field used to trap the ions in the axial direction, this relationship is only approximate. The axial electrical trapping results in axial oscillations within the trap with the (angular) frequency :\omega_\text = \sqrt, where \alpha is a constant similar to the spring constant of a
harmonic oscillator In classical mechanics, a harmonic oscillator is a system that, when displaced from its equilibrium position, experiences a restoring force ''F'' proportional to the displacement ''x'': \vec F = -k \vec x, where ''k'' is a positive const ...
and is dependent on applied voltage, trap dimensions and trap geometry. The electric field and the resulting axial harmonic motion reduces the cyclotron frequency and introduces a second radial motion called magnetron motion that occurs at the magnetron frequency. The cyclotron motion is still the frequency being used, but the relationship above is not exact due to this phenomenon. The natural angular frequencies of motion are :\omega_\pm = \frac \pm \sqrt, where \omega_\text is the axial trapping frequency due the axial electrical trapping and \omega_+ is the reduced cyclotron (angular) frequency and \omega_- is the magnetron (angular) frequency. Again, \omega_+ is what is typically measured in FTICR. The meaning of this equation can be understood qualitatively by considering the case where \omega_\text is much smaller than \omega_\text, which is typically true in the mass spectrometer. In that case, the value of the radical is just slightly less than \omega_\text/2, and the value of \omega_+ is just slightly less than \omega_\text (the cyclotron frequency has been slightly reduced). For \omega_- the value of the radical is the same (slightly less than \omega_\text/2), but it is being subtracted from \omega_\text/2, resulting in a small number equal to \omega_\text - \omega_+ (i.e. the amount that the cyclotron frequency was reduced by). In this regime, the frequencies are approximately \begin \omega_+ &\approx \omega_c - \frac \\ \omega_- &\approx \frac \end


Instrumentation

FTICR-MS differs significantly from other
mass spectrometry Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a ''mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is used ...
techniques in that the ions are not detected by hitting a detector such as an
electron multiplier An electron multiplier is a vacuum-tube structure that multiplies incident charges. In a process called secondary emission, a single electron can, when bombarded on secondary-emissive material, induce emission of roughly 1 to 3 electrons. If an ele ...
but only by passing near detection plates. Additionally the masses are not resolved in space or time as with other techniques but only by the ion cyclotron resonance (rotational) frequency that each ion produces as it rotates in a magnetic field. Thus, the different ions are not detected in different places as with
sector instrument A sector instrument is a general term for a class of mass spectrometer that uses a static electric (E) or magnetic (B) sector or some combination of the two (separately in space) as a mass analyzer. Popular combinations of these sectors have been ...
s or at different times as with
time-of-flight Time of flight (ToF) is the measurement of the time taken by an object, particle or wave (be it acoustic, electromagnetic, etc.) to travel a distance through a medium. This information can then be used to measure velocity or path length, or as a w ...
instruments, but all ions are detected simultaneously during the detection interval. This provides an increase in the observed
signal-to-noise ratio Signal-to-noise ratio (SNR or S/N) is a measure used in science and engineering that compares the level of a desired signal to the level of background noise. SNR is defined as the ratio of signal power to noise power, often expressed in deci ...
owing to the principles of
Fellgett's advantage Fellgett's advantage or the multiplex advantage is an improvement in signal-to-noise ratio (SNR) that is gained when taking multiplexed measurements rather than direct measurements. The name is derived from P. B. Fellgett, who first made the observ ...
. In FTICR-MS, resolution can be improved either by increasing the strength of the magnet (in teslas) or by increasing the detection duration.


Cells

A review of different cell geometries with their specific electric configurations is available in the literature. However, ICR cells can belong to one of the following two categories: closed cells or open cells. Several closed ICR cells with different geometries were fabricated and their performance has been characterized. Grids were used as end caps to apply an axial electric field for trapping ions axially (parallel to the magnetic field lines). Ions can be either generated inside the cell or can be injected to the cell from an external ionization source. Nested ICR cells with double pair of grids were also fabricated to trap both positive and negative ions simultaneously. The most common open cell geometry is a cylinder, which is axially segmented to produce electrodes in the shape of a ring. The central ring electrode is commonly used for applying radial excitation electric field and detection. DC electric voltage is applied on the terminal ring electrodes to trap ions along the magnetic field lines. Open cylindrical cells with ring electrodes of different diameters have also been designed. They proved not only capable in trapping and detecting both ion polarities simultaneously, but also they succeeded to separate positive from negative ions radially. This presented a large discrimination in kinetic ion acceleration between positive and negative ions trapped simultaneously inside the new cell. Several ion axial acceleration schemes were recently written for ion–ion collision studies.


Stored-waveform inverse Fourier transform

Stored-waveform inverse Fourier transform (SWIFT) is a method for the creation of excitation waveforms for FTMS. The time-domain excitation waveform is formed from the inverse Fourier transform of the appropriate frequency-domain excitation spectrum, which is chosen to excite the resonance frequencies of selected ions. The SWIFT procedure can be used to select ions for
tandem mass spectrometry Tandem mass spectrometry, also known as MS/MS or MS2, is a technique in instrumental analysis where two or more stages of analysis using one or more mass analyzer are performed with an additional reaction step in between these analyses to increa ...
experiments.


Applications

Fourier-transform ion cyclotron resonance (FTICR) mass spectrometry is a high-resolution technique that can be used to determine masses with high accuracy. Many applications of FTICR-MS use this mass accuracy to help determine the composition of molecules based on accurate mass. This is possible due to the mass defect of the elements. FTICR-MS is able to achieve higher levels of mass accuracy than other forms of
mass spectrometer Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a '' mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is us ...
, in part, because a superconducting magnet is much more stable than
radio-frequency Radio frequency (RF) is the oscillation rate of an alternating electric current or voltage or of a magnetic, electric or electromagnetic field or mechanical system in the frequency range from around to around . This is roughly between the ...
(RF) voltage. Another place that FTICR-MS is useful is in dealing with complex mixtures, such as biomass or waste liquefaction products, since the resolution (narrow peak width) allows the signals of two ions with similar mass-to-charge ratios (''m''/''z'') to be detected as distinct ions. This high resolution is also useful in studying large macromolecules such as proteins with multiple charges, which can be produced by
electrospray ionization Electrospray ionization (ESI) is a technique used in mass spectrometry to produce ions using an electrospray in which a high voltage is applied to a liquid to create an aerosol. It is especially useful in producing ions from macromolecules becau ...
. For example, attomole level of detection of two peptides has been reported. These large molecules contain a distribution of
isotopes Isotopes are distinct nuclear species (or ''nuclides'') of the same chemical element. They have the same atomic number (number of protons in their nuclei) and position in the periodic table (and hence belong to the same chemical element), but ...
that produce a series of isotopic peaks. Because the isotopic peaks are close to each other on the ''m''/''z'' axis, due to the multiple charges, the high resolving power of the FTICR is extremely useful. FTICR-MS is very useful in other studies of proteomics as well. It achieves exceptional resolution in both top-down and bottom-up proteomics. Electron-capture dissociation (ECD), collisional-induced dissociation (CID), and infrared multiphoton dissociation (IRMPD) are all utilized to produce fragment spectra in tandem mass spectrometry experiments. Although CID and IRMPD use vibrational excitation to further dissociate peptides by breaking the backbone amide linkages, which are typically low in energy and weak, CID and IRMPD may also cause dissociation of post-translational modifications. ECD, on the other hand, allows specific modifications to be preserved. This is quite useful in analyzing phosphorylation states, O- or N-linked glycosylation, and sulfating.


References


External links


What's in an Oil Drop? An Introduction to Fourier Transform Ion Cyclotron Resonance (FT-ICR) for Non-scientists
National High Magnetic Field Laboratory
Scottish Instrumentation Resource Centre for Advanced Mass SpectrometryFourier-transform Ion Cyclotron Resonance (FT-ICR)
FT-ICR Introduction University of Bristol {{DEFAULTSORT:Fourier Transform Ion Cyclotron Resonance Mass spectrometry Measuring instruments