CO oximetry
   HOME

TheInfoList



OR:

A CO-oximeter is a device that measures the
oxygen Oxygen is the chemical element with the symbol O and atomic number 8. It is a member of the chalcogen group in the periodic table, a highly reactive nonmetal, and an oxidizing agent that readily forms oxides with most elements as ...
carrying state of
hemoglobin Hemoglobin (haemoglobin BrE) (from the Greek word αἷμα, ''haîma'' 'blood' + Latin ''globus'' 'ball, sphere' + ''-in'') (), abbreviated Hb or Hgb, is the iron-containing oxygen-transport metalloprotein present in red blood cells (erythrocyt ...
in a
blood Blood is a body fluid in the circulatory system of humans and other vertebrates that delivers necessary substances such as nutrients and oxygen to the cells, and transports metabolic waste products away from those same cells. Blood in the cir ...
specimen, including oxygen-carrying hemoglobin (O2Hb), non-oxygen-carrying but normal hemoglobin (HHb) (formerly, but incorrectly, referred to as 'reduced' hemoglobin), as well as the dyshemoglobins such as carboxyhemoglobin (COHb) and methemoglobin (MetHb). The use of 'CO' rather than 'Co' or 'co' is more appropriate since this designation represents a device that measures
carbon monoxide Carbon monoxide (chemical formula CO) is a colorless, poisonous, odorless, tasteless, flammable gas that is slightly less dense than air. Carbon monoxide consists of one carbon atom and one oxygen atom connected by a triple bond. It is the simple ...
(CO) bound to hemoglobin, as distinguished from simple
oximetry Pulse oximetry is a noninvasive method for monitoring a person's oxygen saturation. Peripheral oxygen saturation (SpO2) readings are typically within 2% accuracy (within 4% accuracy in 95% of cases) of the more accurate (and invasive) reading o ...
which measures hemoglobin bound to molecular oxygen—O2Hb—or hemoglobin capable of binding to molecular oxygen—HHb. Simpler oximeters may report oxygen saturation alone, i.e. the ratio of oxyhemoglobin to total 'bindable' hemoglobin (i.e. oxyhemoglobin + deoxyhemoglobin-HHb). CO-oximetry is useful in defining the causes for
hypoxemia Hypoxemia is an abnormally low level of oxygen in the blood. More specifically, it is oxygen deficiency in arterial blood. Hypoxemia has many causes, and often causes hypoxia as the blood is not supplying enough oxygen to the tissues of the bod ...
, or hypoxia, ( oxygen deficiency at the tissue level).


Mechanism

A CO-oximeter measures the absorption of light passing through blood from few as two or three wavelengths of light to several dozens of
wavelengths In physics, the wavelength is the spatial period of a periodic wave—the distance over which the wave's shape repeats. It is the distance between consecutive corresponding points of the same phase on the wave, such as two adjacent crests, tr ...
, in order to distinguish
oxyhemoglobin Hemoglobin (haemoglobin BrE) (from the Greek word αἷμα, ''haîma'' 'blood' + Latin ''globus'' 'ball, sphere' + ''-in'') (), abbreviated Hb or Hgb, is the iron-containing oxygen-transport metalloprotein present in red blood cells (erythrocyte ...
, and deoxyhemoglobin (formerly called 'reduced' hemoglobin), and thus determine the oxyhemoglobin saturation (the percentage of oxygenated hemoglobin compared to the total amount of available
hemoglobin Hemoglobin (haemoglobin BrE) (from the Greek word αἷμα, ''haîma'' 'blood' + Latin ''globus'' 'ball, sphere' + ''-in'') (), abbreviated Hb or Hgb, is the iron-containing oxygen-transport metalloprotein present in red blood cells (erythrocyt ...
(Hb)). Measurement of greater numbers of wavelengths enables the instrument to distinguish between these and carboxyhemoglobin,-COHb,
methemoglobin Methemoglobin (British: methaemoglobin) (pronounced "met-hemoglobin") is a hemoglobin ''in the form of metalloprotein'', in which the iron in the heme group is in the Fe3+ (ferric) state, not the Fe2+ (ferrous) of normal hemoglobin. Sometimes, it i ...
-metHb, other hemoglobin moieties and 'background' light-absorbing species. Traditionally, measurement is made from
arterial blood Arterial blood is the oxygenated blood in the circulatory system found in the pulmonary vein, the left chambers of the heart, and in the artery, arteries. It is bright red in color, while venous blood is dark red in color (but looks purple through ...
processed in a specific device designed to be able to measure proportions of multiple components of several hemoglobin moieties using multi-wavelength spectrophotometry and complex, but straightforward internal computations. While these units still are in wide use, blood gas analyzers with integral CO-oximetry modules have also been developed and successfully marketed by several manufacturers. More recently, some 'pulse' or more precisely 'peripheral' oximeters have made it possible to estimate carboxyhemoglobin with non-invasive technology similar to a simple (peripheral) pulse oximeter. In contrast, the use of a standard or simple pulse oximeter is not effective in the diagnosis of CO poisoning as patients who have carbon monoxide poisoning may have a normal
oxygen saturation Oxygen saturation (symbol SO2) is a relative measure of the concentration of oxygen that is dissolved or carried in a given medium as a proportion of the maximal concentration that can be dissolved in that medium at the given temperature. It ca ...
reading on a pulse oximeter.


Usage

When a patient presents with
carbon monoxide poisoning Carbon monoxide poisoning typically occurs from breathing in carbon monoxide (CO) at excessive levels. Symptoms are often described as " flu-like" and commonly include headache, dizziness, weakness, vomiting, chest pain, and confusion. Large ...
(CO) or other non-respiratory hypoxic symptoms, most current CO-oximeters will detect the relative levels of each hemoglobin fraction (oxyhemoglobin and dyshemoglobins) and likely the oxyhemoglobin saturation. For any system making these measurements it is critical that the device clearly distinguish between Oxygen Saturation' and Fractional Oxyhemoglobin" . The issue here is the careless use of saturation vs. fractional oxyhemoglobin, which both measure the same entity -oxyhemoglobin- but the oxygen saturation uses as its base only the hemoglobin available for binding, while the fractional oxyhemoglobin uses the total hemoglobin in the sample as its base. In normal subjects the values are nearly identical-thus leading to terminologic and possibly clinical confusion. A simple oximeter measuring only oxygen derivatives, may report a normal saturation or even a hyperoxic state if oxygen gas has been administered when in fact there is serious compromise of oxygen carrying ability of the hemoglobin present.


See also

*
Breath carbon monoxide Breath carbon monoxide is the level of carbon monoxide in a person's exhalation. It can be measured in a breath carbon monoxide test, generally by using a carbon monoxide breath monitor (breath CO monitor), such as for motivation and education for ...


References

{{reflist *CLSI, C46-A2- Blood Gas and pH Analysis and Related Measurements; Approved Guideline—Second Edition, Wayne, PA, 2010 *Zijlstra WJ, Maas AHJ, Moran RF. Definition, significance and measurement of quantities pertaining to the oxygen carrying properties of human blood.. ''Scand J Clin Lab Invest'' , 56(Suppl), 224, 27–45, 1996 *Brunelle JA, Degtiarov AM, Moran RF, Race LA, Simultaneous measurement of total hemoglobin and its derivatives in blood using CO-oximeters: Analytical principles; Their application in selecting analytical wavelengths and reference methods; A comparison of the results of the choices made. ''Scand J Clin Lab Invest,'' 56: (Suppl) 224, 47–69, 1996. *Brunelle JA, Moran RF, Data processing in CO-oximeters that Use overdetermined systems (Reply). ''Clin Chem,'' 43:1, 189–191, 1997 *Degen BR, Moran RF, Comparison and assessment of blood gas related quantities including base excess, the gas exchange indices and temperature corrected pH/ ''P''O2/''P''CO2 , as defined in approved NCCLS standard C12-A, using a computer simulation of input variables., ''Scand J Clin Lab Invest ,'' 56:(Suppl) 224, 89-106 1996. *Moran R, Hemoglobin F and measurement of oxygen saturation and fractional oxyhemoglobin. ''Clin Lab Sci,'' 7:3, 162–164, 1994. *Brunnelle JA, Degtiarov AM, Moran RF, Race LA, CO-oximetric measurement of oxyhemoglobin, deoxyhemoglobin, and dyshemoglobins in blood: Effects of analytical wavelength and reference method selection. ''Lab Hematol''. 1:2, 161 - 164, 1995. *Moran RF, Implications of Fetal Hemoglobin : Measurement of oxygen saturation, fractional oxyhemoglobin, carboxyhemoglobin and methemoglobin. ''Crit Care International'', April–May, 8–9, 1995. *Moran RF, The case for standardized terminology: Oxygen "saturation" values can trick the unwary and lead to clinical misjudgement., ''Crit Care Med'', 21:5, 805–807, 1993. *Moran RF, Lab Consultant: igh Percentage of zero carboxyhemoglobins due to correction algorithm for small, "impossible", values.''Clin Chem News'', 18:12, 18–19, 1992. *CLSI document C25A, can provide in-depth information and references. Medical testing equipment Oxygen de:Pulsoxymetrie#CO-Oxymeter