Definition
CIPM definition of the sievert
The SI definition given by the International Committee for Weights and Measures (CIPM) says: "The quantity dose equivalent ''H'' is the product of the absorbed dose ''D'' of ionizing radiation and the dimensionless factor ''Q'' (quality factor) defined as a function of linear energy transfer by the ICRU" :''H'' = ''Q'' × ''D'' The value of ''Q'' is not defined further by CIPM, but it requires the use of the relevant ICRU recommendations to provide this value. The CIPM also says that "in order to avoid any risk of confusion between the absorbed dose ''D'' and the dose equivalent ''H'', the special names for the respective units should be used, that is, the name gray should be used instead of joules per kilogram for the unit of absorbed dose ''D'' and the name sievert instead of joules per kilogram for the unit of dose equivalent ''H''". In summary: :gray: quantity ''D'' – absorbed dose ::1 Gy = 1 joule/kilogram – a physical quantity. 1 Gy is the deposit of a joule of radiation energy per kilogram of matter or tissue. :sievert: quantity ''H'' – dose equivalent ::1 Sv = 1 joule/kilogram – a biological effect. The sievert represents the equivalent biological effect of the deposit of a joule of radiation energy in a kilogram of human tissue. The ratio to absorbed dose is denoted by ''Q''.ICRP definition of the sievert
The ICRP definition of the sievert is:ICRP publication 103 - Glossary. : "The sievert is the special name for the SI unit of equivalent dose, effective dose, and operational dose quantities. The unit is joule per kilogram." The sievert is used for a number of dose quantities which are described in this article and are part of the international radiological protection system devised and defined by the ICRP and ICRU.External dose quantities
When the sievert is used to represent the stochastic effects of external ionizing radiation on human tissue, the radiation doses received are measured in practice by radiometric instruments and dosimeters and are called operational quantities. To relate these actual received doses to likely health effects, protection quantities have been developed to predict the likely health effects using the results of large epidemiological studies. Consequently, this has required the creation of a number of different dose quantities within a coherent system developed by the ICRU working with the ICRP. The external dose quantities and their relationships are shown in the accompanying diagram. The ICRU is primarily responsible for the operational dose quantities, based upon the application of ionising radiation metrology, and the ICRP is primarily responsible for the protection quantities, based upon modelling of dose uptake and biological sensitivity of the human body.Naming conventions
The ICRU/ICRP dose quantities have specific purposes and meanings, but some use common words in a different order. There can be confusion between, for instance, ''equivalent dose'' and ''dose equivalent''. Although the CIPM definition states that the linear energy transfer function (Q) of the ICRU is used in calculating the biological effect, the ICRP in 1990 developed the "protection" dose quantities ''effective'' and ''equivalent'' dose which are calculated from more complex computational models and are distinguished by not having the phrase ''dose equivalent'' in their name. Only the operational dose quantities which still use Q for calculation retain the phrase ''dose equivalent''. However, there are joint ICRU/ICRP proposals to simplify this system by changes to the operational dose definitions to harmonise with those of protection quantities. These were outlined at the 3rd International Symposium on Radiological Protection in October 2015, and if implemented would make the naming of operational quantities more logical by introducing "dose to lens of eye" and "dose to local skin" as ''equivalent doses''. In the USA there are differently named dose quantities which are not part of the ICRP nomenclature.Physical quantities
These are directly measurable physical quantities in which no allowance has been made for biological effects. Radiation fluence is the number of radiation particles impinging per unit area per unit time, kerma is the ionising effect on air ofOperational quantities
Operational quantities are measured in practice, and are the means of directly measuring dose uptake due to exposure, or predicting dose uptake in a measured environment. In this way they are used for practical dose control, by providing an estimate or upper limit for the value of the protection quantities related to an exposure. They are also used in practical regulations and guidance. The calibration of individual and area dosimeters in photon fields is performed by measuring the collision "air kerma free in air" under conditions of secondary electron equilibrium. Then the appropriate operational quantity is derived applying a conversion coefficient that relates the air kerma to the appropriate operational quantity. The conversion coefficients for photon radiation are published by the ICRU. Simple (non-anthropomorphic) "phantoms" are used to relate operational quantities to measured free-air irradiation. The ICRU sphere phantom is based on the definition of an ICRU 4-element tissue-equivalent material which does not really exist and cannot be fabricated. The ICRU sphere is a theoretical 30 cm diameter "tissue equivalent" sphere consisting of a material with a density of 1 g·cm−3 and a mass composition of 76.2% oxygen, 11.1% carbon, 10.1% hydrogen and 2.6% nitrogen. This material is specified to most closely approximate human tissue in its absorption properties. According to the ICRP, the ICRU "sphere phantom" in most cases adequately approximates the human body as regards the scattering and attenuation of penetrating radiation fields under consideration. Thus radiation of a particular energy fluence will have roughly the same energy deposition within the sphere as it would in the equivalent mass of human tissue. To allow for back-scattering and absorption of the human body, the "slab phantom" is used to represent the human torso for practical calibration of whole body dosimeters. The slab phantom is depth to represent the human torso. The joint ICRU/ICRP proposals outlined at the 3rd International Symposium on Radiological Protection in October 2015 to change the definition of operational quantities would not change the present use of calibration phantoms or reference radiation fields.Protection quantities
Protection quantities are calculated models, and are used as "limiting quantities" to specify exposure limits to ensure, in the words of ICRP, "that the occurrence of stochastic health effects is kept below unacceptable levels and that tissue reactions are avoided". These quantities cannot be measured in practice but their values are derived using models of external dose to internal organs of the human body, using anthropomorphic phantoms. These are 3D computational models of the body which take into account a number of complex effects such as body self-shielding and internal scattering of radiation. The calculation starts with organ absorbed dose, and then applies radiation and tissue weighting factors. As protection quantities cannot practically be measured, operational quantities must be used to relate them to practical radiation instrument and dosimeter responses.Instrument and dosimetry response
This is an actual reading obtained from such as an ambient dose gamma monitor, or a personal dosimeter. Such instruments are calibrated using radiation metrology techniques which will trace them to a national radiation standard, and thereby relate them to an operational quantity. The readings of instruments and dosimeters are used to prevent the uptake of excessive dose and to provide records of dose uptake to satisfy radiation safety legislation; such as in the UK, theCalculating protection dose quantities
The sievert is used in external radiation protection for equivalent dose (the external-source, whole-body exposure effects, in a uniform field), and effective dose (which depends on the body parts irradiated). These dose quantities are weighted averages of absorbed dose designed to be representative of the stochastic health effects of radiation, and use of the sievert implies that appropriate weighting factors have been applied to the absorbed dose measurement or calculation (expressed in grays). The ICRP calculation provides two weighting factors to enable the calculation of protection quantities. : 1. The radiation factor ''W''''R'', which is specific for radiation type ''R'' – This is used in calculating the equivalent dose ''H''''T'' which can be for the whole body or for individual organs. : 2. The tissue weighting factor ''W''''T'', which is specific for tissue type T being irradiated. This is used with ''W''''R'' to calculate the contributory organ doses to arrive at an effective dose ''E'' for non-uniform irradiation. When a whole body is irradiated uniformly only the radiation weighting factor ''W''''R'' is used, and the effective dose equals the whole body equivalent dose. But if the irradiation of a body is partial or non-uniform the tissue factor ''W''''T'' is used to calculate dose to each organ or tissue. These are then summed to obtain the effective dose. In the case of uniform irradiation of the human body, these summate to 1, but in the case of partial or non-uniform irradiation, they will summate to a lower value depending on the organs concerned; reflecting the lower overall health effect. The calculation process is shown on the accompanying diagram. This approach calculates the biological risk contribution to the whole body, taking into account complete or partial irradiation, and the radiation type or types. The values of these weighting factors are conservatively chosen to be greater than the bulk of experimental values observed for the most sensitive cell types, based on averages of those obtained for the human population.Radiation type weighting factor ''W''''R''
Since different radiation types have different biological effects for the same deposited energy, a corrective radiation weighting factor ''WR'', which is dependent on the radiation type and on the target tissue, is applied to convert the absorbed dose measured in the unit gray to determine the equivalent dose. The result is given the unit sievert. The equivalent dose is calculated by multiplying the absorbed energy, averaged by mass over an organ or tissue of interest, by a radiation weighting factor appropriate to the type and energy of radiation. To obtain the equivalent dose for a mix of radiation types and energies, a sum is taken over all types of radiation energy dose. where : is the equivalent dose absorbed by tissue ''T'', : is the absorbed dose in tissue ''T'' by radiation type ''R'' and : is the radiation weighting factor defined by regulation. Thus for example, an absorbed dose of 1 Gy by alpha particles will lead to an equivalent dose of 20 Sv. This may seem to be a paradox. It implies that the energy of the incident radiation field in joules has increased by a factor of 20, thereby violating the laws ofTissue type weighting factor ''W''''T''
The second weighting factor is the tissue factor ''W''''T'', but it is used only if there has been non-uniform irradiation of a body. If the body has been subject to uniform irradiation, the effective dose equals the whole body equivalent dose, and only the radiation weighting factor ''W''''R'' is used. But if there is partial or non-uniform body irradiation the calculation must take account of the individual organ doses received, because the sensitivity of each organ to irradiation depends on their tissue type. This summed dose from only those organs concerned gives the effective dose for the whole body. The tissue weighting factor is used to calculate those individual organ dose contributions. The ICRP values for ''W''''T'' are given in the table shown here. The article on effective dose gives the method of calculation. The absorbed dose is first corrected for the radiation type to give the equivalent dose, and then corrected for the tissue receiving the radiation. Some tissues like bone marrow are particularly sensitive to radiation, so they are given a weighting factor that is disproportionally large relative to the fraction of body mass they represent. Other tissues like the hard bone surface are particularly insensitive to radiation and are assigned a disproportionally low weighting factor. In summary, the sum of tissue-weighted doses to each irradiated organ or tissue of the body adds up to the effective dose for the body. The use of effective dose enables comparisons of overall dose received regardless of the extent of body irradiation.Operational quantities
The operational quantities are used in practical applications for monitoring and investigating external exposure situations. They are defined for practical operational measurements and assessment of doses in the body. Three external operational dose quantities were devised to relate operational dosimeter and instrument measurements to the calculated protection quantities. Also devised were two phantoms, The ICRU "slab" and "sphere" phantoms which relate these quantities to incident radiation quantities using the Q(L) calculation.Ambient dose equivalent
This is used for area monitoring of penetrating radiation and is usually expressed as the quantity ''H''*(10). This means the radiation is equivalent to that found 10 mm within the ICRU sphere phantom in the direction of origin of the field. An example of penetrating radiation is gamma rays.Directional dose equivalent
This is used for monitoring of low penetrating radiation and is usually expressed as the quantity ''H(0.07). This means the radiation is equivalent to that found at a depth of 0.07 mm in the ICRU sphere phantom. Examples of low penetrating radiation are alpha particles, beta particles and low-energy photons. This dose quantity is used for the determination of equivalent dose to such as the skin, lens of the eye. In radiological protection practice value of omega is usually not specified as the dose is usually at a maximum at the point of interest.Personal dose equivalent
This is used for individual dose monitoring, such as with a personal dosimeter worn on the body. The recommended depth for assessment is 10 mm which gives the quantity ''H''p(10).Proposals for changing the definition of protection dose quantities
In order to simplify the means of calculating operational quantities and assist in the comprehension of radiation dose protection quantities, ICRP Committee 2 & ICRU Report Committee 26 started in 2010 an examination of different means of achieving this by dose coefficients related to Effective Dose or Absorbed Dose. Specifically; 1. For area monitoring of effective dose of whole body it would be: :''H'' = Φ × conversion coefficient The driver for this is that ''H''∗(10) is not a reasonable estimate of effective dose due to high energy photons, as a result of the extension of particle types and energy ranges to be considered in ICRP report 116. This change would remove the need for the ICRU sphere and introduce a new quantity called ''E''max. 2. For individual monitoring, to measure deterministic effects on eye lens and skin, it would be: :''D'' = Φ × conversion coefficient for absorbed dose. The driver for this is the need to measure the deterministic effect, which it is suggested, is more appropriate than stochastic effect. This would calculate equivalent dose quantities ''H''lens and ''H''skin. This would remove the need for the ICRU Sphere and the Q-L function. Any changes would replace ICRU report 51, and part of report 57."Operational Quantities and new approach by ICRU" – Akira Endo. The 3rd International Symposium on the System of Radiological Protection, Seoul, Korea – October 20–22, 201Internal dose quantities
The sievert is used for human internal dose quantities in calculatingHealth effects
Ionizing radiation has deterministic and stochastic effects on human health. Deterministic (acute tissue effect) events happen with certainty, with the resulting health conditions occurring in every individual who received the same high dose. Stochastic (cancer induction and genetic) events are inherentlyStochastic effects
Stochastic effects are those that occur randomly, such as radiation-induced cancer. The consensus of nuclear regulators, governments and theDeterministic effects
The deterministic (acute tissue damage) effects that can lead to acute radiation syndrome only occur in the case of acute high doses (≳ 0.1 Gy) and high dose rates (≳ 0.1 Gy/h) and are conventionally not measured using the unit sievert, but use the unit gray (Gy). A model of deterministic risk would require different weighting factors (not yet established) than are used in the calculation of equivalent and effective dose.ICRP dose limits
The ICRP recommends a number of limits for dose uptake in table 8 of report 103. These limits are "situational", for planned, emergency and existing situations. Within these situations, limits are given for the following groups: * Planned exposure – limits given for occupational, medical and public * Emergency exposure – limits given for occupational and public exposure * Existing exposure – All persons exposed For occupational exposure, the limit is 50 mSv in a single year with a maximum of 100 mSv in a consecutive five-year period, and for the public to an average of 1 mSv (0.001 Sv) of effective dose per year, not including medical and occupational exposures. For comparison, natural radiation levels inside theDose examples
Significant radiation doses are not frequently encountered in everyday life. The following examples can help illustrate relative magnitudes; these are meant to be examples only, not a comprehensive list of possible radiation doses. An "acute dose" is one that occurs over a short and finite period of time, while a "chronic dose" is a dose that continues for an extended period of time so that it is better described by a dose rate.Dose examples
Dose rate examples
All conversions between hours and years have assumed continuous presence in a steady field, disregarding known fluctuations, intermittent exposure and radioactive decay. Converted values are shown in parentheses. "/a" is "per annum", which means per year. "/h" means "per hour". Notes on examples:History
The sievert has its origin in theCommon SI usage
Frequently used SI prefixes are the millisievert (1 mSv = 0.001 Sv) and microsievert (1 μSv = 0.000 001 Sv) and commonly used units for time derivative or "dose rate" indications on instruments and warnings for radiological protection are μSv/h and mSv/h. Regulatory limits and chronic doses are often given in units of mSv/a or Sv/a, where they are understood to represent an average over the entire year. In many occupational scenarios, the hourly dose rate might fluctuate to levels thousands of times higher for a brief period of time, without infringing on the annual limits. The conversion from hours to years varies because of leap years and exposure schedules, but approximate conversions are: : 1 mSv/h = 8.766 Sv/a : 114.1 μSv/h = 1 Sv/a Conversion from hourly rates to annual rates is further complicated by seasonal fluctuations in natural radiation, decay of artificial sources, and intermittent proximity between humans and sources. The ICRP once adopted fixed conversion for occupational exposure, although these have not appeared in recent documents: : 8 h = 1 day : 40 h = 1 week : 50 weeks = 1 year Therefore, for occupation exposures of that time period, : 1 mSv/h = 2 Sv/a : 500 μSv/h = 1 Sv/aIonizing radiation quantities
The following table shows radiation quantities in SI and non-SI units: Although the United States Nuclear Regulatory Commission permits the use of the unitsRem equivalence
An older unit for the dose equivalent is theSee also