The Spectronic 20 is a
brand A brand is a name, term, design, symbol or any other feature that distinguishes one seller's good or service from those of other sellers. Brands are used in business, marketing, and advertising for recognition and, importantly, to create ...
of single-beam spectrophotometer, designed to operate in the
visible spectrum The visible spectrum is the portion of the electromagnetic spectrum that is visible to the human eye. Electromagnetic radiation in this range of wavelengths is called ''visible light'' or simply light. A typical human eye will respond to wa ...
across a wavelength range of 340 nm to 950 nm, with a spectral bandpass of 20 nm. It is designed for
quantitative Quantitative may refer to: * Quantitative research, scientific investigation of quantitative properties * Quantitative analysis (disambiguation) * Quantitative verse, a metrical system in poetry * Statistics, also known as quantitative analysis ...
absorption measurement at single wavelengths. Because it measures the
transmittance Transmittance of the surface of a material is its effectiveness in transmitting radiant energy. It is the fraction of incident electromagnetic power that is transmitted through a sample, in contrast to the transmission coefficient, which is t ...
or absorption of visible light through a solution, it is sometimes referred to as a colorimeter. The name of the instrument is a
trademark A trademark (also written trade mark or trade-mark) is a type of intellectual property consisting of a recognizable sign, design, or expression that identifies products or services from a particular source and distinguishes them from othe ...
of the manufacturer. Developed by
Bausch & Lomb Bausch + Lomb is an eye health products company based in Vaughan, Ontario, Canada. It is one of the world's largest suppliers of contact lenses, lens care products, pharmaceuticals, intraocular lenses, and other eye surgery products. The compan ...
and launched in 1953, the Spectronic 20 was the first low-cost spectrophotometer. It rapidly became an industry standard due to its low cost, durability and ease of use, and has been referred to as an "iconic lab spectrophotometer". Approximately 600,000 units were sold over its nearly 60 year production run. It has been the most widely used spectrophotometer worldwide. Production was discontinued in 2011 when it was replaced by the Spectronic 200,, but the Spectronic 20 is still in common use. It is sometimes referred to as the "Spec 20".


Design

The Bausch & Lomb Spectronic 20 colorimeter uses a
diffraction grating In optics, a diffraction grating is an optical component with a periodic structure that diffracts light into several beams travelling in different directions (i.e., different diffraction angles). The emerging coloration is a form of structur ...
monochromator A monochromator is an optical device that transmits a mechanically selectable narrow band of wavelengths of light or other radiation chosen from a wider range of wavelengths available at the input. The name is from the Greek roots ''mono-'', "s ...
combined with a system for the detection, amplification, and measurement of light wavelengths in the 340 nm to 950 nm range. As shown in the schematic optical diagram (see left), polychromatic light from a source in the system passes through lenses which are reflected and dispersed by the diffraction grating to restrict the range of light wavelengths. This restricted range of wavelengths is then passed through the sample to be measured. The intensity of the transmitted light is determined by a
phototube A phototube or photoelectric cell is a type of gas-filled or vacuum tube that is sensitive to light. Such a tube is more correctly called a 'photoemissive cell' to distinguish it from photovoltaic or photoconductive cells. Phototubes were previ ...
detector A sensor is a device that produces an output signal for the purpose of sensing a physical phenomenon. In the broadest definition, a sensor is a device, module, machine, or subsystem that detects events or changes in its environment and sends ...
. Mechanical movement of the diffraction grating by means of the cam attached to the wavelength control enables the user to select for various wavelengths. This is the "λ knob", wherein λ refers to wavelength of light used for the measurement.


Quantitative measurements

Many substances absorb light in the ultraviolet - visible light range. Absorption at any particular wavelength in the ultraviolet visible range is proportional to the concentration of the substances in the solution or other medium, in accord with the Beer–Lambert relationship. In a practical sense, the Beer–Lambert relationship can be stated as: ''A'' = ε x ''l'' x ''c'' in which ''A'' is the absorbance measured by the instrument, ε is the molar absorption coefficient of the sample, ''l'' is the pathlength of the light beam through the sample, and c is the concentration of the substance in the solution or medium. The Spectronic 20 is thereby commonly used for quantitative determination of the concentration of a substance of interest. The Spectronic 20 measures the absorbance of light at a pre-determined concentration, and the concentration is calculated from the Beer–Lambert relationship. The absorbance of the light is the base 10 logarithm of the
ratio In mathematics, a ratio shows how many times one number contains another. For example, if there are eight oranges and six lemons in a bowl of fruit, then the ratio of oranges to lemons is eight to six (that is, 8:6, which is equivalent to the ...
of the
Transmittance Transmittance of the surface of a material is its effectiveness in transmitting radiant energy. It is the fraction of incident electromagnetic power that is transmitted through a sample, in contrast to the transmission coefficient, which is t ...
of the pure solvent to the transmittance of the sample, and so the two absorbance and transmittance can be interconverted. Either transmittance or absorbance can therefore be plotted versus concentration using measurements from the Spectronic 20. Plotting a curve using percent transmittance of light yields an exponential curve. However, absorbance is linearly related to concentration, and so absorbance is often preferred for plotting a standard curve. This type of standard curve relates the concentration of the solution (on the x-axis) to measures of its absorbance (y-axis). To obtain such a curve, a series of dilutions of known concentration of a solution are prepared and readings are obtained for each of the dilutions (see plot at left). In this plot, the slope of the line is the product ε x ''l''. By measuring a series of standards and creating the standard curve, it is possible to quantify the amount or concentration of a substance within a sample by determining the absorbance on the Spec 20 and finding the corresponding concentration on the calibration curve. Alternatively, the logarithm of percent transmittance can be plotted versus concentration to create a standard curve using the same procedure. The absorbance measured by the Spectronic 20 is the sum of the absorbance of each of the constituents of the solution. Therefore, the Spectronic 20 can be used to analyze more complex solutions. For example, if a sample solution has two light-absorbing compounds in it, then the user performs measurements at two different wavelengths and constructs standard curves for each compound. Then the concentration of each compound can be calculated algebraically.Wikipedia contributors. "
Beer–Lambert law The Beer–Lambert law, also known as Beer's law, the Lambert–Beer law, or the Beer–Lambert–Bouguer law relates the attenuation of light to the properties of the material through which the light is travelling. The law is commonly applied t ...
." Wikipedia, The Free Encyclopedia. Wikipedia, The Free Encyclopedia, 15 Mar. 2017. Web. 8 Apr. 2017.
The Spectronic 20 can be used for turbidimetric measurements. In microbiological work, the turbidity of a liquid culture of bacterial cells relates to the cell count, and OD600 measurements can be conducted for this purpose using the Spectronic 20. Likewise the turbidity of water suspensions of
clays Clay is a type of fine-grained natural soil material containing clay minerals (hydrous aluminium phyllosilicates, e.g. kaolin, Al2 Si2 O5( OH)4). Clays develop plasticity when wet, due to a molecular film of water surrounding the clay par ...
and other
particles In the physical sciences, a particle (or corpuscule in older texts) is a small localized object which can be described by several physical or chemical properties, such as volume, density, or mass. They vary greatly in size or quantity, from s ...
of size suitable for
light scattering Scattering is a term used in physics to describe a wide range of physical processes where moving particles or radiation of some form, such as light or sound, are forced to deviate from a straight trajectory by localized non-uniformities (including ...
can be quantitatively determined by means of a Spectronic 20. In the past, the Spectronic 20 was used for clinical diagnostic purposes.


Use

Before testing a sample, the Spectronic 20 is calibrated using a blank solution, which is the pure solvent that is used in the experimental sample. It is typically water or an
organic solvent A solvent (s) (from the Latin '' solvō'', "loosen, untie, solve") is a substance that dissolves a solute, resulting in a solution. A solvent is usually a liquid but can also be a solid, a gas, or a supercritical fluid. Water is a solvent for ...
. In this calibration, the transmittance is set at 100% using the calibration knob of the instrument (the amplifier control knob in the figure at right). The instrument can also optionally be calibrated with a stock solution of a sample at a concentration known to have an absorbance of 2 or else vendor supplied standards, using the light absorption knob in the diagram shown at right. After calibration, the user places a 1/2 inch test tube or
cuvette A cuvette (French: cuvette = "little vessel") is a small tube-like container with straight sides and a circular or square cross section. It is sealed at one end, and made of a clear, transparent material such as plastic, glass, or fused quartz. ...
containing the sample solution to be measured into the sample compartment. Calibration is repeated each time the wavelength is changed. It or a standard reference sample is generally used to periodically check for
drift Drift or Drifts may refer to: Geography * Drift or ford (crossing) of a river * Drift, Kentucky, unincorporated community in the United States * In Cornwall, England: ** Drift, Cornwall, village ** Drift Reservoir, associated with the village ...
. To measure wavelengths above 650 nm, the bottom of the instrument is opened, and a red
filter Filter, filtering or filters may refer to: Science and technology Computing * Filter (higher-order function), in functional programming * Filter (software), a computer program to process a data stream * Filter (video), a software component tha ...
and a red-sensitive
photocell Photodetectors, also called photosensors, are sensors of light or other electromagnetic radiation. There is a wide variety of photodetectors which may be classified by mechanism of detection, such as photoelectric or photochemical effects, or b ...
is installed. The original design of the Spectronic 20 utilized an
analog Analog or analogue may refer to: Computing and electronics * Analog signal, in which information is encoded in a continuous variable ** Analog device, an apparatus that operates on analog signals *** Analog electronics, circuits which use analog ...
dial for readout of transmission from 100%T to 1%T (top scale), 0A - 2A (lower scale). Using the original instrument requires manual setting of the
wavelength In physics, the wavelength is the spatial period of a periodic wave—the distance over which the wave's shape repeats. It is the distance between consecutive corresponding points of the same phase on the wave, such as two adjacent crests, t ...
and making readings from a moving-needle analog display.


Replacement

The Spectronic 20D (launched in 1985) and later the 20D+ replaced the analog dial with a red digital LED readout, offering greater
precision Precision, precise or precisely may refer to: Science, and technology, and mathematics Mathematics and computing (general) * Accuracy and precision, measurement deviation from true value and its scatter * Significant figures, the number of digit ...
in the readout, if not greater
accuracy Accuracy and precision are two measures of ''observational error''. ''Accuracy'' is how close a given set of measurements ( observations or readings) are to their ''true value'', while ''precision'' is how close the measurements are to each oth ...
in the actual reading. A side-by-side comparison of the features of the 20+ and 20D+ is available in the 2001 operating manual. The Spectronic 20 was replaced by the Spectronic 200 in the Thermo Scientific spectrophotometer product line in 2011. The Spectronic 200 utilizes an array detector and digital control of the measured wavelength, while retaining the characteristic λ knob of the Spec 20 for setting the wavelength. In addition to replicating the user modes of the Spec 20D+ (which it can emulate on a color LCD screen) the Spec 200 accommodates both test-tubes and square cuvettes without needing to install an adapter. Software modes described in the Spectronic 200's specifications include scanning, four wavelength simultaneous measurement, and quantitative analysis with up to four standards, in contrast to the SPEC 20D+ which offered only single point calibration.


Product line history

Originally introduced by Bausch & Lomb in 1953, the product line was sold to Milton Roy in 1985. Milton Roy sold its instrument group to Life Sciences International, renamed Spectronic instruments, Inc. in 1995. Spectronics Instruments was purchased by Thermo Optek in 1997, renamed Spectronic-Unicam in 2001 and Thermo-Spectronic in 2002. In 2003 the product line was moved to
Madison, WI Madison is the county seat of Dane County and the capital city of the U.S. state of Wisconsin. As of the 2020 census the population was 269,840, making it the second-largest city in Wisconsin by population, after Milwaukee, and the 80th-larg ...
and the brand renamed to
Thermo Electron Thermo Electron Corporation (NYSE: TMO) (incorporated 1956) was a major provider of analytical instruments and services for a variety of domains. It was founded in 1956 by George N. Hatsopoulos, an MIT PhD in mechanical engineering. Initial fu ...
. With the merger of Thermo Electron and
Fisher Scientific Fisher Scientific International, Inc. (NYSE: FSH) was a laboratory supply and biotechnology company that provided products and services to the global scientific research and clinical laboratory markets until its merger with Thermo Electron in 200 ...
in 2006 the brand changed to Thermo Scientific, and remained such until the end of the production run. Spectronic 20 instruments found in labs today may bear any of the Bausch and Lomb, Milton Roy, Spectronic, Thermo Electron or Thermo Scientific brand names.


Popular culture

The Spectronic 20 is apparently one of the few lab instruments to remain intact after the destruction of the laboratory in the movie ''