Fast Low Angle Shot Magnetic Resonance Imaging
   HOME

TheInfoList



OR:

Fast low angle shot magnetic resonance imaging (FLASH MRI) is a particular sequence of
magnetic resonance imaging Magnetic resonance imaging (MRI) is a medical imaging technique used in radiology to form pictures of the anatomy and the physiological processes of the body. MRI scanners use strong magnetic fields, magnetic field gradients, and radio wave ...
. It is a gradient echo sequence which combines a low-flip angle radio-frequency excitation of the nuclear magnetic resonance signal (recorded as a spatially encoded gradient echo) with a short repetition time. It is the generic form of steady-state free precession imaging. Different manufacturers of MRI equipment use different names for this experiment.
Siemens Siemens AG ( ) is a German multinational conglomerate corporation and the largest industrial manufacturing company in Europe headquartered in Munich with branch offices abroad. The principal divisions of the corporation are ''Industry'', '' ...
uses the name FLASH, General Electric used the name SPGR (Spoiled Gradient Echo), and Philips uses the name CE-FFE-T1 (Contrast-Enhanced Fast Field Echo) or T1-FFE. Depending on the desired contrast, the generic FLASH technique provides spoiled versions that destroy transverse coherences and yield T1 contrast as well as refocused versions (constant phase per repetition) and fully balanced versions (zero phase per repetition) that incorporate transverse coherences into the steady-state signal and offer T1/T2 contrast. Applications include: *cross-sectional images with acquisition times of a few seconds enable MRI studies of the thorax and abdomen within a single breathhold, *dynamic acquisitions synchronized to the electrocardiogram generate movies of the beating heart, *sequential acquisitions monitor physiological processes such as the differential uptake of contrast media into body tissues, * three-dimensional acquisitions visualize complex anatomic structures (brain, joints) at unprecedented high spatial resolution in all three dimensions and along arbitrary view directions, and *
Magnetic resonance angiography Magnetic resonance angiography (MRA) is a group of techniques based on magnetic resonance imaging (MRI) to image blood vessels. Magnetic resonance angiography is used to generate images of arteries (and less commonly veins) in order to evaluate t ...
(MRA) yields three-dimensional representations of the vasculature.


Physical basis

The physical basis of MRI is the spatial encoding of the nuclear magnetic resonance (NMR) signal obtainable from water protons (i.e. hydrogen nuclei) in biologic tissue. In terms of MRI, signals with different spatial encodings that are required for the reconstruction of a full image need to be acquired by generating multiple signals – usually in a repetitive way using multiple radio-frequency excitations. The generic FLASH technique emerges as a gradient echo sequence which combines a low-flip angle radio-frequency excitation of the NMR signal (recorded as a spatially encoded gradient echo) with a rapid repetition of the basic sequence. The repetition time is usually much shorter than the typical T1 relaxation time of the protons in biologic tissue. Only the combination of (i) a low-flip angle excitation which leaves unused longitudinal magnetization for an immediate next excitation with (ii) the acquisition of a gradient echo which does not need a further radio-frequency pulse that would affect the residual longitudinal magnetization, allows for the rapid repetition of the basic sequence interval and the resulting speed of the entire image acquisition. In fact, the FLASH sequence eliminated all waiting periods previously included to accommodate effects from T1 saturation. FLASH reduced the typical sequence interval to what is minimally required for imaging: a slice-selective radio-frequency pulse and gradient, a phase-encoding gradient, and a (reversed) frequency-encoding gradient generating the echo for data acquisition. For radial data sampling, the phase- and frequency-encoding gradients are replaced by two simultaneously applied frequency-encoding gradients that rotate the Fourier lines in data space. In either case, repetition times are as short as 2 to 10 milliseconds, so that the use of 64 to 256 repetitions results in image acquisition times of about 0.1 to 2.5 seconds for a two-dimensional image. Most recently, highly undersampled radial FLASH MRI acquisitions have been combined with an iterative image reconstruction by regularized nonlinear inversion to achieve real-time MRI at a temporal resolution of 20 to 30 milliseconds for images with a spatial resolution of 1.5 to 2.0 millimeters. This method allows for a visualization of the beating heart in real time – without synchronization to the electrocardiogram and during free breathing.


History

FLASH MRI was invented in 1985 by
Jens Frahm Jens Frahm (born 29 March 1951 in Oldenburg, Germany) is a German biophysicist and physicochemist. He is Research Group Leader of the Biomedical NMR group at the Max Planck Institute (MPI) for Multidisciplinary Sciences in Göttingen, Germany ( ...
, Axel Haase, W Hänicke, KD Merboldt, and D Matthaei (German Patent Application P 35 04 734.8, 12 February 1985) at th
Max-Planck-Institut für biophysikalische Chemie
in Göttingen, Germany. The technique is revolutionary in shortening MRI measuring times by up to two orders of magnitude. FLASH was very rapidly adopted commercially. RARE was slower, and echo-planar imaging (EPI) – for technical reasons – took even more time. Echo-planar imaging had been proposed by Mansfield's group in 1977, and the first crude images were shown by Mansfield and Ian Pykett in the same year. Roger Ordidge presented the first movie in 1981. Its breakthrough came with the invention of shielded gradients.A SHORT HISTORY OF MAGNETIC RESONANCE IMAGING FROM A EUROPEAN POINT OF VIEW
/ref> The introduction of FLASH MRI sequences in diagnostic imaging for the first time allowed for a drastic shortening of the measuring times without a substantial loss in image quality. In addition, the measuring principle led to a broad range of completely new imaging modalities. In 2010, an extended FLASH method with highly undersampled radial data encoding and iterative image reconstruction achieved real-time MRI with a temporal resolution of 20
millisecond A millisecond (from '' milli-'' and second; symbol: ms) is a unit of time in the International System of Units (SI) equal to one thousandth (0.001 or 10−3 or 1/1000) of a second and to 1000 microseconds. A unit of 10 milliseconds may be called ...
s (1/50th of a second). Taken together, this latest development corresponds to an acceleration by a factor of 10,000 compared to the MRI situation before 1985. In general, FLASH denoted a breakthrough in clinical MRI that stimulated further technical as well as scientific developments up to date.


References

{{Reflist, refs= {{ cite patent , country=DE , number=35 04 734.8 , status=patent , title=Hochfrequenz-Impuls und Gradienten-Impuls-Verfahren zur Aufnahme von schnellen NMR-Tomogrammen unter Benutzung von Gradientenechos , pubdate=1986-08-14 , gdate=1998-12-10 , fdate=1985-02-12 , inventor= J Frahm, A Haase, W Hänicke, KD Merboldt, D Matthaei {{cite journal , author= J Frahm, A Haase, D Matthaei , date=1986 , title=Rapid NMR imaging of dynamic processes using the FLASH technique , journal=Magnetic Resonance in Medicine , volume=3 , issue=2 , pages=321–327 , doi=10.1002/mrm.1910030217 , pmid=3713496 , s2cid=31028542 {{cite journal , author= J Frahm, A Haase, D Matthaei , date=1986 , title=Rapid three-dimensional MR imaging using the FLASH technique , journal=Journal of Computer Assisted Tomography , volume=10 , issue=2 , pages=363–368 , pmid=3950172 , doi=10.1097/00004728-198603000-00046 {{cite journal , author=A Haase, J Frahm, D Matthaei, W Hänicke, KD Merboldt , date=1986 , title=FLASH imaging: rapid NMR imaging using low flip angle pulses , journal=Journal of Magnetic Resonance , volume=67 , issue=2 , pages=258–266 , doi=10.1016/0022-2364(86)90433-6 , bibcode=1986JMagR..67..258H {{cite journal , author= D Matthaei, J Frahm, A Haase, W Hänicke , date=1985 , title=Regional physiological functions depicted by sequences of rapid magnetic resonance images , journal= The Lancet , volume=326 , issue=8460 , pages=893 , doi=10.1016/S0140-6736(85)90158-8 , pmid=2864605 , s2cid=12326347 {{cite journal , author=M Uecker, S Zhang, D Voit, A Karaus, KD Merboldt, J Frahm , date=2010 , title=Real-time MRI at a resolution of 20 ms , journal= NMR in Biomedicine , volume=23 , issue=8 , pages=986–994 , doi=10.1002/nbm.1585 , pmid=20799371 , hdl=11858/00-001M-0000-0012-D4F9-7 , s2cid=8268489 , hdl-access=free {{cite journal , author=S Zhang, M Uecker, D Voit, KD Merboldt, J Frahm , date=2010 , title=Real-time cardiovascular magnetic resonance at high temporal resolution: radial FLASH with nonlinear inverse reconstruction , journal=Journal of Cardiovascular Magnetic Resonance , volume=12 , pages=39 , doi=10.1186/1532-429X-12-39 , pmid=20615228 , pmc=2911425 {{cite journal , author=S Zhang, KT Block, J Frahm , date=2010 , title=Magnetic resonance imaging in real time: Advances using radial FLASH , journal=Journal of Magnetic Resonance Imaging , volume=31 , issue=1 , pages=101–109 , doi=10.1002/jmri.21987 , pmid=19938046 , hdl=11858/00-001M-0000-0012-D667-0 , s2cid=17419027 , hdl-access=free


External links


Biomedizinische NMR Forschungs GmbH
offers further detailed information about FLASH MRI and related applications (neurobiology, cardiovascular imaging)

of the
Max Planck Society The Max Planck Society for the Advancement of Science (german: Max-Planck-Gesellschaft zur Förderung der Wissenschaften e. V.; abbreviated MPG) is a formally independent non-governmental and non-profit association of German research institutes. ...
* http://www.mtbeurope.info/news/2010/1009005.htm Magnetic resonance imaging