In electron energy loss spectroscopy (EELS) a material is exposed to a beam of
electron The electron ( or ) is a subatomic particle with a negative one elementary electric charge. Electrons belong to the first generation of the lepton particle family, and are generally thought to be elementary particles because they have n ...
s with a known, narrow range of kinetic energies. Some of the electrons will undergo inelastic scattering, which means that they lose energy and have their paths slightly and randomly deflected. The amount of energy loss can be measured via an electron spectrometer and interpreted in terms of what caused the energy loss. Inelastic interactions include
phonon In physics, a phonon is a collective excitation in a periodic, elastic arrangement of atoms or molecules in condensed matter, specifically in solids and some liquids. A type of quasiparticle, a phonon is an excited state in the quantum mechani ...
excitations, inter- and intra-band transitions,
plasmon In physics, a plasmon is a quantum of plasma oscillation. Just as light (an optical oscillation) consists of photons, the plasma oscillation consists of plasmons. The plasmon can be considered as a quasiparticle since it arises from the quantiz ...
excitations, inner shell
ionization Ionization, or Ionisation is the process by which an atom or a molecule acquires a negative or positive charge by gaining or losing electrons, often in conjunction with other chemical changes. The resulting electrically charged atom or molecul ...
s, and Cherenkov radiation. The inner-shell ionizations are particularly useful for detecting the elemental components of a material. For example, one might find that a larger-than-expected number of electrons comes through the material with 285  eV less energy than they had when they entered the material. This is approximately the amount of energy needed to remove an inner-shell electron from a carbon atom, which can be taken as evidence that there is a significant amount of
carbon Carbon () is a chemical element with the symbol C and atomic number 6. It is nonmetallic and tetravalent—its atom making four electrons available to form covalent chemical bonds. It belongs to group 14 of the periodic table. Carbon ma ...
present in the sample. With some care, and looking at a wide range of energy losses, one can determine the types of atoms, and the numbers of atoms of each type, being struck by the beam. The scattering angle (that is, the amount that the electron's path is deflected) can also be measured, giving information about the
dispersion relation In the physical sciences and electrical engineering, dispersion relations describe the effect of dispersion on the properties of waves in a medium. A dispersion relation relates the wavelength or wavenumber of a wave to its frequency. Given t ...
of whatever material excitation caused the inelastic scattering.


History

The technique was developed by James Hillier and RF Baker in the mid-1940s but was not widely used over the next 50 years, only becoming more widespread in research in the 1990s due to advances in microscope instrumentation and vacuum technology. With modern instrumentation becoming widely available in laboratories worldwide, the technical and scientific developments from the mid-1990s have been rapid. The technique is able to take advantage of modern aberration-corrected probe forming systems to attain spatial resolutions down to ~0.1 nm, while with a monochromated electron source and/or careful deconvolution the energy resolution can be 0.1 eV or better. This has enabled detailed measurements of the atomic and electronic properties of single columns of atoms, and in a few cases, of single atoms.


Comparison with EDX

EELS is spoken of as being complementary to
energy-dispersive x-ray spectroscopy Energy-dispersive X-ray spectroscopy (EDS, EDX, EDXS or XEDS), sometimes called energy dispersive X-ray analysis (EDXA or EDAX) or energy dispersive X-ray microanalysis (EDXMA), is an analytical technique used for the elemental analysis or chemi ...
(variously called EDX, EDS, XEDS, etc.), which is another common spectroscopy technique available on many electron microscopes. EDX excels at identifying the atomic composition of a material, is quite easy to use, and is particularly sensitive to heavier elements. EELS has historically been a more difficult technique but is in principle capable of measuring atomic composition, chemical bonding, valence and conduction band electronic properties, surface properties, and element-specific pair distance distribution functions. EELS tends to work best at relatively low atomic numbers, where the excitation edges tend to be sharp, well-defined, and at experimentally accessible energy losses (the signal being very weak beyond about 3 keV energy loss). EELS is perhaps best developed for the elements ranging from carbon through the 3d transition metals (from
scandium Scandium is a chemical element with the symbol Sc and atomic number 21. It is a silvery-white metallic d-block element. Historically, it has been classified as a rare-earth element, together with yttrium and the Lanthanides. It was discovered in ...
to
zinc Zinc is a chemical element with the symbol Zn and atomic number 30. Zinc is a slightly brittle metal at room temperature and has a shiny-greyish appearance when oxidation is removed. It is the first element in group 12 (IIB) of the periodi ...
). For carbon, an experienced spectroscopist can tell at a glance the differences between diamond, graphite, amorphous carbon, and "mineral" carbon (such as the carbon appearing in carbonates). The spectra of 3d transition metals can be analyzed to identify the oxidation states of the atoms. Cu(I), for instance, has a different so-called "white-line" intensity ratio than Cu(II) does. This ability to "fingerprint" different forms of the same element is a strong advantage of EELS over EDX. The difference is mainly due to the difference in energy resolution between the two techniques (~1 eV or better for EELS, perhaps a few tens of eV for EDX).


Variants

There are several basic flavors of EELS, primarily classified by the geometry and by the kinetic energy of the incident electrons (typically measured in kiloelectron-volts, or keV). Probably the most common today is transmission EELS, in which the kinetic energies are typically 100 to 300 keV and the incident electrons pass entirely through the material sample. Usually this occurs in a transmission electron microscope (TEM), although some dedicated systems exist which enable extreme resolution in terms of energy and momentum transfer at the expense of spatial resolution. Other flavors include reflection EELS (including reflection high-energy electron energy-loss spectroscopy (RHEELS)), typically at 10 to 30 keV, and aloof EELS (sometimes called near-field EELS), in which the electron beam does not in fact strike the sample but instead interacts with it via the long-ranged Coulomb interaction. Aloof EELS is particularly sensitive to surface properties but is limited to very small energy losses such as those associated with surface plasmons or direct interband transitions. Within transmission EELS, the technique is further subdivided into valence EELS (which measures plasmons and interband transitions) and inner-shell ionization EELS (which provides much the same information as
x-ray absorption spectroscopy X-ray absorption spectroscopy (XAS) is a widely used technique for determining the local geometric and/or electronic structure of matter. The experiment is usually performed at synchrotron radiation facilities, which provide intense and tunabl ...
, but from much smaller volumes of material). The dividing line between the two, while somewhat ill-defined, is in the vicinity of 50 eV energy loss. Instrumental developments have opened up the ultra-low energy loss part of the EELS spectrum, enabling
vibrational spectroscopy Infrared spectroscopy (IR spectroscopy or vibrational spectroscopy) is the measurement of the interaction of infrared radiation with matter by absorption, emission, or reflection. It is used to study and identify chemical substances or function ...
in the TEM. Both IR-active and non-IR-active vibrational modes are present in EELS.


EEL spectrum

The electron energy loss (EEL) spectrum can be roughly split into two different regions: the low-loss spectrum (up until about 50eV in energy loss) and the high-loss spectrum. The low-loss spectrum contains the zero-loss peak as well as the plasmon peaks, and contains information about the band structure and dielectric properties of the sample. The high-loss spectrum contains the ionisation edges that arise due to inner shell ionisations in the sample. These are characteristic to the species present in the sample, and as such can be used to obtain accurate information about the chemistry of a sample.


Thickness measurements

EELS allows quick and reliable measurement of local thickness in
transmission electron microscopy Transmission electron microscopy (TEM) is a microscopy technique in which a beam of electrons is transmitted through a specimen to form an image. The specimen is most often an ultrathin section less than 100 nm thick or a suspension on a ...
. The most efficient procedure is the following: * Measure the energy loss spectrum in the energy range about −5..200 eV (wider better). Such measurement is quick (milliseconds) and thus can be applied to materials normally unstable under electron beams. * Analyse the spectrum: (i) extract zero-loss peak (ZLP) using standard routines; (ii) calculate integrals under the ZLP (''I0'') and under the whole spectrum (''I''). * The thickness ''t'' is calculated as mfp*''ln(I/I0)''. Here mfp is the mean free path of electron inelastic scattering, which has been tabulated for most elemental solids and oxides. The spatial resolution of this procedure is limited by the plasmon localization and is about 1 nm, meaning that spatial thickness maps can be measured in
scanning transmission electron microscopy A scanning transmission electron microscope (STEM) is a type of transmission electron microscope (TEM). Pronunciation is tɛmor �sti:i:ɛm As with a conventional transmission electron microscope (CTEM), images are formed by electrons passin ...
with ~1 nm resolution.


Pressure measurements

The intensity and position of low-energy EELS peaks are affected by pressure. This fact allows mapping local pressure with ~1 nm spatial resolution. * ''Peak shift method'' is reliable and straightforward. The peak position is calibrated by independent (usually optical) measurement using a
diamond anvil cell A diamond anvil cell (DAC) is a high-pressure device used in geology, engineering, and materials science experiments. It enables the compression of a small (sub-millimeter-sized) piece of material to extreme pressures, typically up to around 1 ...
. However, the spectral resolution of most EEL spectrometers (0.3-2 eV, typically 1 eV) is often too crude for the small pressure-induced shifts. Therefore, the sensitivity and accuracy of this method is relatively poor. Nevertheless, pressures as small as 0.2 GPa inside helium bubbles in aluminum have been measured. * ''Peak intensity method'' relies on pressure-induced change in the intensity of dipole-forbidden transitions. Because this intensity is zero for zero pressure the method is relatively sensitive and accurate. However, it requires existence of allowed and forbidden transitions of similar energies and thus is only applicable to specific systems, e.g., Xe bubbles in aluminum.


Use in confocal geometry

Scanning confocal electron energy loss microscopy (SCEELM) is a new analytical microscopy tool that enables a double corrected transmission electron microscope to achieve sub-10 nm depth resolution in depth sectioning imaging of nanomaterials. It was previously termed as energy filtered scanning confocal electron microscopy due to the lack to full spectrum acquisition capability (only a small energy window on the order of 5 eV can be used at a time). SCEELM takes advantages of the newly developed chromatic aberration corrector which allows electrons of more than 100 eV of energy spread to be focused to roughly the same focal plane. It has been demonstrated that a simultaneous acquisition of the zero loss, low-loss, and core loss signals up to 400 eV in the confocal geometry with depth discrimination capability.


See also

* Energy filtered transmission electron microscopy * Magic angle (EELS) *
Transmission electron microscopy Transmission electron microscopy (TEM) is a microscopy technique in which a beam of electrons is transmitted through a specimen to form an image. The specimen is most often an ultrathin section less than 100 nm thick or a suspension on a ...
*
Scanning transmission electron microscopy A scanning transmission electron microscope (STEM) is a type of transmission electron microscope (TEM). Pronunciation is tɛmor �sti:i:ɛm As with a conventional transmission electron microscope (CTEM), images are formed by electrons passin ...


References


Further reading

* * * *


External links


A Database of EELS fine structure fingerprints at Cornell

A database of EELS and X-Ray excitation spectra

Cornell Spectrum Imager, an EELS Analysis open-source plugin for ImageJ

HyperSpy, a hyperspectral data analysis Python toolbox especially well suited for EELS data analysis

EELSMODEL, software to quantify Electron Energy Loss (EELS) spectra by using model fitting
{{Branches of Spectroscopy Electron spectroscopy Scientific techniques