A design history file is a compilation of
document A document is a written, drawn, presented, or memorialized representation of thought, often the manifestation of non-fictional, as well as fictional, content. The word originates from the Latin ''Documentum'', which denotes a "teaching" o ...
ation that describes the design history of a finished
medical device A medical device is any device intended to be used for medical purposes. Significant potential for hazards are inherent when using a device for medical purposes and thus medical devices must be proved safe and effective with reasonable assura ...
. The design history file, or DHF, is part of regulation introduced in 1990 when the
U.S. Congress The United States Congress is the legislature of the federal government of the United States. It is bicameral, composed of a lower body, the House of Representatives, and an upper body, the Senate. It meets in the U.S. Capitol in Washin ...
passed the Safe Medical Devices Act, which established new standards for medical devices that can cause or contribute to the death, serious illness, or injury of a
patient A patient is any recipient of health care services that are performed by healthcare professionals. The patient is most often ill or injured and in need of treatment by a physician, nurse, optometrist, dentist, veterinarian, or other hea ...
. Prior to this legislation,
U.S. Food and Drug Administration The United States Food and Drug Administration (FDA or US FDA) is a federal agency of the Department of Health and Human Services. The FDA is responsible for protecting and promoting public health through the control and supervision of food ...
(FDA) auditors were limited to examining the production and quality control records of the device.


Requirements

The regulation requires
medical device A medical device is any device intended to be used for medical purposes. Significant potential for hazards are inherent when using a device for medical purposes and thus medical devices must be proved safe and effective with reasonable assura ...
manufacturers of Class II and Class III devices to implement design controls. These design controls consist of a development and control plan used to manage the development of a new product, and a design history file where these activities are documented. These controls are specifically intended to manage a medical device company's
new product development In business and engineering, new product development (NPD) covers the complete process of bringing a new product to market, renewing an existing product or introducing a product in a new market. A central aspect of NPD is product design, along ...
activities. Research and development processes aimed at developing new underlying technologies are not subject to these regulations. The requirements for a DHF are documented in FDA Regulation CFR 21 820.


Design controls

Each manufacturer of either a class II or class III medical device (as well as a select group of class I devices) needs to establish and document procedures on the design and design requirements. These design controls include: Also available in hardcopy, as *Design input - Design inputs are typically the initial requirements that describe the medical device to be produced. *Design output - Design outputs are the results of the design and engineering efforts. These are normally the final specifications for the device. Including the manufacturing process and the in-coming, in-process and finished device inspection, measurement or test methods and criteria. The outputs are normally documented in models, drawings, engineering analysis and other documents. The output needs to be directly traceable to the input requirements. Design verification and validation should demonstrate that the final output specifications conform to the input requirements and meet user needs and intended uses. *Design review - The design review is a formal review of the medical device design by representatives of each design function participating in the design efforts as well as other interested parties (e.g. marketing, sales, manufacturing engineering, etc.). The design review must be documented in the DHF and include review date, participants, design version/revision reviewed and review results. *Design verification - Design verification is the process that confirms that the design output conforms to the design input. Design verification should demonstrate that the specifications are the correct specifications for the design. Design verification must be documented in the DHF and include the verification date, participants, design version/revision verified, verification method and verification results. *Process validation - Process validation is the process in which the device design is validated using initial/low volume production processes. The purpose for the process validation is to confirm that the design functions according to design inputs when produced using normal production processes rather than prototype processes. The process validation must be documented in the DHF. *Design validation. Each manufacturer shall establish and maintain procedures for validating the device design. Design validation shall be performed under defined operating conditions on initial production units, lots, or batches, or their equivalents. Design validation shall ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions. Design validation shall include software validation and risk analysis, where appropriate. The results of the design validation, including identification of the design, method(s), the date, and the individual(s) performing the validation, shall be documented in the DHF. *Design transfer - Design transfer is the process in which the device design is translated into production, distribution, and installation specifications. *Design changes - Design changes is the process in which the design changes are identified and documented. Also known as engineering change or enterprise change. *Design history file - The DHF is a formal document that is prepared for each medical device. The DHF can be either a collection of the actual documents generated in the product development (PD) process or an index of documents and their storage location.


Design and development files

The sub-clause 7.3.10 of ISO 13485:2016 requires a manufacturer of a
medical device A medical device is any device intended to be used for medical purposes. Significant potential for hazards are inherent when using a device for medical purposes and thus medical devices must be proved safe and effective with reasonable assura ...
to maintain (and control) a ''design and development file'' for a medical device to document the design history of a medical device. This file shall also contain records for changes in design and development (per device type or family). It might contain e.g. a design and development plan, or test reports; and is thus comparable to the DHF of the FDA regulations. Similarly, Annex II ยง3.1 of the
EU medical device regulation Regulation (EU) 2017/745 is a regulation of the European Union on the clinical investigation and sale of medical devices for human use. It repeals Directive 93/42/EEC (MDD), which concerns medical devices, and Directive 90/385/EEC, which concer ...
asks for ''information to allow the design stages applied to the device to be understood'' to be part of the
Technical documentation Technical documentation is a generic term for the classes of information created to describe (in technical language) the use, functionality or architecture of a product, system or service. Classes of technical documentation Classes of technical do ...
.


See also

*
Device Master Record A Device Master Record (DMR) is a compilation of all the instructions, drawings and other records that must be used to produce a product. The term is used in Quality Management Systems that cover product design and production. Contents The physica ...