A radiotherapy accident in Columbus, Ohio, also known as the Riverside radiation case, occurred as the result of an incorrectly calibrated cobalt teletherapy unit, occurred between 1974 and 1976, leading to 10 deaths.


Background

Between 1958 and 1972, the Riverside Methodist Hospital in Columbus, Ohio became the first hospital in
Central Ohio According to the U.S. Census Bureau, the Columbus, Ohio Metropolitan Statistical Area includes the counties of Delaware, Fairfield, Franklin, Hocking, Licking, Madison, Morrow, Perry, Pickaway, and Union. The population of the MSA is 2,078,725 acc ...
to develop an extensive cobalt therapy program, where the use of cobalt-60 became the dominant radiation source for treating patients with cancer. In 1973, 30-year-old Joel Axt was hired by the hospital as the resident physicist, as part of a plan to further expand the hospital's radiation therapy program. Axt was previously a teacher at the Xavier University of Louisiana, and had clinical experience at the University of California Medical Center, though his experience was limited to 14 months and was not enough to qualify for
American Board of Radiology Established in 1934, the American Board of Radiology (ABR) is an independent, not-for-profit professional association with headquarters in Tucson, Arizona. It oversees the certification and ongoing professional development of physician specialists ...
certification. After Axt's arrival, he had to "reconstruct Riverside’s radiation physics program almost from scratch," as the previous contracted physicists had removed all his equipment.


Incident

In 1974, Axt calibrated a cobalt-60 teletherapy unit with an incorrect decay curve. The calculated
decay rate Radioactive decay (also known as nuclear decay, radioactivity, radioactive disintegration, or nuclear disintegration) is the process by which an unstable atomic nucleus loses energy by radiation. A material containing unstable nuclei is consid ...
was more rapid than the real source, leading to a dose rate underestimated by 10–45% and an overestimated treatment time. No other calibrations and checks were performed between May 1974 and January 1976. Axt later attributed it to his other demanding responsibilities and high-priority projects at the hospital. By January 1976, patients had been complaining about what would be symptoms of radiation overexposure. A radiation therapist prompted Axt to make a measurement of the output, revealing Axt's mistake, and the teletherapy unit was adjusted. In March, the hospital asked an external team from the MD Anderson Cancer Center to review the accident. After the accident was leaked to the public in April, the
Nuclear Regulatory Commission The Nuclear Regulatory Commission (NRC) is an independent agency of the United States government tasked with protecting public health and safety related to nuclear energy. Established by the Energy Reorganization Act of 1974, the NRC began operat ...
began its own investigation. During the investigations, Axt attributed the high output measured to a faulty measurement system, and produced ten calibration documents to support this theory, though under further investigations he admitted that he had falsified his reports. Nine out of ten of Axt's reports were found to be fabricated. Over a 22-month period, 426 patients received significant overdoses. Around three hundred of the patients died within a year, mainly due to their pre-existing cancer. Of the 183 patients who survived the first year, 88 showed "immediate severe complications related to the irradiated sites." 10 fatalities were known. As a result of this accident, the Nuclear Regulatory Commission issued extensive regulations on the training requirements, and quality assurance procedures required when using cobalt-60 machines. 102 lawsuits have been filed by the survivors and families of the deceased. Axt was fired and disappeared, but he was later found in Miami, and took a deposition in 1977.