Capillary electrochromatography (CEC) is a
chromatographic In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent (gas or liquid) called the ''mobile phase'', which carries it through a system (a ...
technique in which the mobile phase is driven through the chromatographic bed by electroosmosis. Capillary electrochromatography is a combination of two analytical techniques,
high-performance liquid chromatography High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify each component in a mixture. It relies on pumps to pa ...
and
capillary electrophoresis Capillary electrophoresis (CE) is a family of electrokinetic separation methods performed in submillimeter diameter capillaries and in micro- and nanofluidic channels. Very often, CE refers to capillary zone electrophoresis (CZE), but other electr ...
. Capillary electrophoresis aims to separate analytes on the basis of their mass-to-charge ratio by passing a high voltage across ends of a
capillary tube Capillary action (sometimes called capillarity, capillary motion, capillary rise, capillary effect, or wicking) is the process of a liquid flowing in a narrow space without the assistance of, or even in opposition to, any external forces li ...
, which is filled with the analyte. High-performance liquid chromatography separates analytes by passing them, under high pressure, through a column filled with stationary phase. The interactions between the analytes and the stationary phase and
mobile phase In analytical and organic chemistry, elution is the process of extracting one material from another by washing with a solvent; as in washing of loaded ion-exchange resins to remove captured ions. In a liquid chromatography experiment, for exa ...
lead to the separation of the analytes. In capillary electrochromatography capillaries, packed with HPLC stationary phase, are subjected to a high voltage. Separation is achieved by electrophoretic migration of solutes and differential partitioning.


Principle

Capillary electrochromatography (CEC) combines the principles used in HPLC and CE. The mobile phase is driven across the chromatographic bed using electroosmosis instead of pressure (as in HPLC). Electroosmosis is the motion of liquid induced by an applied potential across a porous material, capillary tube, membrane or any other fluid conduit. Electroosmotic flow is caused by the Coulomb force induced by an electric field on net mobile electric charge in a solution. Under
alkaline In chemistry, an alkali (; from ar, القلوي, al-qaly, lit=ashes of the saltwort) is a base (chemistry), basic, ionic compound, ionic salt (chemistry), salt of an alkali metal or an alkaline earth metal. An alkali can also be defined as ...
conditions, the surface
silanol A silanol is a functional group in silicon chemistry with the connectivity Si–O–H. It is related to the hydroxy functional group (C–O–H) found in all alcohols. Silanols are often invoked as intermediates in organosilicon ch ...
groups of the fused silica will become ionised leading to a negatively charged surface. This surface will have a layer of positively charged ions in close proximity which are relatively immobilised. This layer of ions is called the
Stern layer The stern is the back or aft-most part of a ship or boat, technically defined as the area built up over the sternpost, extending upwards from the counter rail to the taffrail. The stern lies opposite the bow, the foremost part of a ship. Or ...
. The thickness of the double layer is given by the formula: :\delta=\sqrt where εr is the
relative permittivity The relative permittivity (in older texts, dielectric constant) is the permittivity of a material expressed as a ratio with the electric permittivity of a vacuum. A dielectric is an insulating material, and the dielectric constant of an insul ...
of the medium, εo is the permittivity of vacuum, R is the
universal gas constant The molar gas constant (also known as the gas constant, universal gas constant, or ideal gas constant) is denoted by the symbol or . It is the molar equivalent to the Boltzmann constant, expressed in units of energy per temperature increment per ...
, T is the
absolute temperature Thermodynamic temperature is a quantity defined in thermodynamics as distinct from kinetic theory or statistical mechanics. Historically, thermodynamic temperature was defined by Kelvin in terms of a macroscopic relation between thermodynamic wor ...
, c is the
molar concentration Molar concentration (also called molarity, amount concentration or substance concentration) is a measure of the concentration of a chemical species, in particular of a solute in a solution, in terms of amount of substance per unit volume of solut ...
, and F is the
Faraday constant In physical chemistry, the Faraday constant, denoted by the symbol and sometimes stylized as ℱ, is the electric charge per mole of elementary charges. It is named after the English scientist Michael Faraday. Since the 2019 redefinition of S ...
When an electric field is applied to the fluid (usually via electrodes placed at inlets and outlets), the net charge in the electrical double layer is induced to move by the resulting Coulomb force. The resulting flow is termed electroosmotic flow. In CEC positive ions of the electrolyte added along with the analyte accumulate in the electrical double layer of the particles of the column packing on application of an electric field they move towards the cathode and drag the liquid mobile phase with them. The relationship between the linear velocity u of the liquid in the capillary and the applied electric field is given by the
Smoluchowski equation Marian Smoluchowski (; 28 May 1872 – 5 September 1917) was a Polish physicist who worked in the Polish territories of the Austro-Hungarian Empire. He was a pioneer of statistical physics, and an avid mountaineer. Life Born into an upper- ...
as : u = \epsilon_r\epsilon_0 \zeta E \eta where ζ is the potential across the Stern layer (
zeta potential Zeta potential is the electrical potential at the slipping plane. This plane is the interface which separates mobile fluid from fluid that remains attached to the surface. Zeta potential is a scientific term for electrokinetic potential in coll ...
), E is the
electric field strength An electric field (sometimes E-field) is the physical field that surrounds electrically charged particles and exerts force on all other charged particles in the field, either attracting or repelling them. It also refers to the physical field f ...
, and η is the
viscosity The viscosity of a fluid is a measure of its resistance to deformation at a given rate. For liquids, it corresponds to the informal concept of "thickness": for example, syrup has a higher viscosity than water. Viscosity quantifies the inte ...
of the solvent. Separation of components in CEC is based on interactions between the stationary phase and differential electrophoretic migration of solutes.


Instrumentation

The components of a capillary electrochromatograph are a sample
vial A vial (also known as a phial or flacon) is a small glass or plastic vessel or bottle, often used to store medication as liquids, powders or capsules. They can also be used as scientific sample vessels; for instance, in autosampler devices i ...
, source and destination vials, a packed capillary, electrodes, a
high voltage power supply A power supply is an electrical device that supplies electric power to an electrical load. The main purpose of a power supply is to convert electric current from a source to the correct voltage, current, and frequency to power the load. As a r ...
, a detector, and a data output and handling device. The source vial, destination vial and capillary are filled with an electrolyte such as an aqueous buffer solution. The capillary is packed with stationary phase. To introduce the sample, the capillary inlet is placed into a vial containing the sample and then returned to the source vial (sample is introduced into the capillary via capillary action, pressure, or siphoning). The migration of the analytes is then initiated by an electric field that is applied between the source and destination vials and is supplied to the electrodes by the high-voltage power supply. The analytes separate as they migrate due to their electrophoretic mobility, and are detected near the outlet end of the capillary. The output of the detector is sent to a data output and handling device such as an integrator or computer. The data is then displayed as an
electropherogram An electropherogram, or electrophoregram, can also be referred to as an EPG or e-gram. It is a record or chart produced when electrophoresis is used in an analytical technique, primarily in the fields of forensic biology, molecular biology and ...
, which reports detector response as a function of time. Separated chemical compounds appear as peaks with different migration times in an
electropherogram An electropherogram, or electrophoregram, can also be referred to as an EPG or e-gram. It is a record or chart produced when electrophoresis is used in an analytical technique, primarily in the fields of forensic biology, molecular biology and ...
.


Advantages

Avoiding the use of pressure to introduce the mobile phase into the column, results in a number of important advantages. Firstly, the pressure driven flow rate across a column depends directly on the square of the particle diameter and inversely on the length of the column. This restricts the length of the column and size of the particle, particle size is seldom less than 3 micrometer and the length of the column is restricted to 25 cm. Electrically driven flow rate is independent of length of column and size. A second advantage of using electroosmosis to pass the mobile phase into the column is the plug-like flow velocity profile of EOF, which reduces the solute dispersion in the column, increasing column efficiency.


See also

*
Chromatography In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent (gas or liquid) called the ''mobile phase'', which carries it through a system (a ...
* Electrophoresis *
High-performance liquid chromatography High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify each component in a mixture. It relies on pumps to pa ...
*
Capillary electrophoresis Capillary electrophoresis (CE) is a family of electrokinetic separation methods performed in submillimeter diameter capillaries and in micro- and nanofluidic channels. Very often, CE refers to capillary zone electrophoresis (CZE), but other electr ...
*
Electrochromatography Electrochromatography is a chemical separation technique in analytical chemistry, biochemistry and molecular biology used to resolve and separate mostly large biomolecules such as proteins. It is a combination of size exclusion chromatography (gel ...


References


Further reading

* Smith, N. "Capillary ElectroChromatography" Available at:https://www.beckmancoulter.com/wsrportal/bibliography?docname=AP8508ACECPrimer.pdf * Bartle, K. D. ''Capillary ElectroChromatography'' Published by The Royal Society of Chemistry, Cambridge. {{Electrophoresis Chromatography Protein methods Molecular biology Laboratory techniques Electrophoresis Biological techniques and tools